User's Manual
Performance Tests
AWG2005 Service Manual
4Ć39
b. Select waveform file:
Push SETUPWaveform Sequence, if necessary, to select a
waveform file for CH1. Waveform Sequence toggles between the
CH1 files (upper list) and the CH2 files (lower list).
Turn the general purpose knob to highlight the CNTRL.WFM
file.
Push ENTER to select the file.
3. Check rearĆpanel CONTROL SIG OUT amplitude:
a. Check CONTROL SIG OUT pulse amplitude:
Push ModeGated.
Push the AWG2005 MANUAL TRIGGER button.
Check that the pulse upper level of the displayed waveform is
2ĂV or greater.
Check that the pulse lower level of the displayed waveform is
0.8ĂV or less.
4. End procedure: Disconnect the oscilloscope.
This procedure checks the external trigger level accuracy of the AWG2005.
Electrical Characteristic Checked: Auxiliary Input, TRIGGER, Accuracy,
on pageĂ1Ć15.
Equipment Required: Two 50ĂΩ coaxial cables, a function generator, and
an oscilloscope.
Prerequisites: The AWG2005 meets the prerequisites listed on page 4Ć7.
Procedure:
1. Install test hookup and set test equipment controls:
a. Hook up oscilloscope: Connect the AWG2005 CH1 output through a
coaxial cable to the oscilloscope CH1 vertical input.
b. Hook up function generator: Connect the AWG2005 TRIGGER INPUT
through a coaxial cable to the function generator output (see FigĆ
ure 4Ć18).
External Trigger Level
Accuracy Check