User's Manual

Performance Tests
Performance Verification
4Ć38
This procedure checks the amplitude of the CONTROL SIG OUT signal.
Electrical Characteristic Checked: Auxiliary Output, CONTROL SIG,
Amplitude, on page 1Ć14Ă.
Equipment Required: A 50Ă coaxial cable and an oscilloscope.
Prerequisites: The AWG2005 meets the prerequisites listed on page 4Ć7.
Procedure:
1.        
a.     Connect the AWG2005 rearĆpanel CONĆ
TROL SIG OUT connector through the coaxial cable to the CH1
vertical input connector on the oscilloscope (see Figure 4Ć17).
OscilloscopeAWG2005
AWG2005
Figure 4Ć17:ăInitial Test Hookup
b.   
Vertical CH1
CH1 Coupling DC
CH1 Scale 1ĂV/div.
CH1 Input Impedance 50Ă
Horizontal
Sweep 50Ăns/div.
Trigger
Source CH1
Coupling DC
Slope Positive
Level 500ĂmV
Mode Auto
2.         
a.     Push UTILITYMiscConfig...ReĆ
set to FactoryO.K.
CONTROL SIG OUT
Amplitude Check