User's Manual

Performance Tests
AWG2005 Service Manual
4Ć37
Vertical CH1
CH1 Coupling DC
CH1 Scale 500ĂmV/div.
CH1 Input Impedance 50Ă
Horizontal
Sweep 50õs/div.
Trigger
Source CH1
Coupling DC
Slope Positive
Level 500ĂmV
Mode Auto
2. Set the AWG2005 controls and select the waveform file:
a. Initialize AWG2005 controls: Push UTILITYMiscConfig...ReĆ
set to FactoryO.K.
b. Select waveform file:
Push SETUPWaveform Sequence, if necessary, to select a
waveform file for CH1. Waveform Sequence toggles between the
CH1 files (upper list) and the CH2 files (lower list).
Turn the general purpose knob to highlight the MRK.WFM file.
Push ENTER to select the file.
3. Check rearĆpanel CH1 MARKER OUT amplitude:
a. Check CH1 MARKER OUT pulse amplitude:
Check that the pulse amplitude of the displayed waveform is
2ĂV
pĆp
or greater.
4. Check rearĆpanel CH2 MARKER OUT pulse amplitude:
a. Check CH2 MARKER OUT pulse amplitude:
Move the coaxial cable from the AWG2005 rearĆpanel CH1
MARKER OUT connector to the rearĆpanel CH2 MARKER OUT
connector.
Check that the pulse amplitude of the displayed waveform is 2ĂV
or greater.
5. Check Option 02: If the AWG2005 has CH3 and CH4, repeat this entire
test, selecting the AWG2005 waveform and setting controls for CH3 and
CH4 and checking:
RearĆpanel CH3 MARKER pulse amplitude
RearĆpanel CH4 MARKER pulse amplitude
6. End procedure: Disconnect the oscilloscope.