User's Manual

Performance Tests
AWG2005 Service Manual
4Ć35
2. Set the AWG2005 controls and select the waveform file:
a. Initialize AWG2005 controls: Push UTILITYMiscConfig...ReĆ
set to FactoryO.K.
b. Select waveform file:
Push SETUPWaveform Sequence, if necessary, to select a
waveform file for CH1. Waveform Sequence toggles between the
CH1 files (upper list) and the CH2 files (lower list).
Turn the general purpose knob to select the PULSE.WFM file.
Push ENTER to select the file.
Select the PULSE.WFM file for CH2 same as CH1.
3. Turn on the AWG2005 CH1 output: Push the CH1 button so that the LED
above the CH1 output connector is on.
4. Check pulse response at 0.5ĂV amplitude:
a. Check rise time: Check that the rise time of the waveform displayed
on the oscilloscope from the 10% point to the 90% point is 35Ăns or
less.
b. Check aberrations: Check that the aberrations of the displayed
waveform is within 0.35Ădiv.
c. Check flatness: Check that the flatness of the displayed waveform is
within 0.15Ădiv. after 20Ăns from the rising edge.
d. Change the oscilloscope controls:
Trigger
Slope Negative
e. Check fall time: Check that the fall time of the displayed waveform
from the 10% point to the 90% point is 35Ăns or less.
5. Check pulse response at 10ĂV amplitude:
a. Change the oscilloscope controls:
Vertical CH1
CH1 scale 2ĂV/div.
Trigger
Slope Positive
b. Change the AWG2005 controls:
Push SETUPAmplitude to change the amplitude for CH1.
Press the numeric key 1, 0 and press the units key V to select
an amplitude of 10ĂV.
c. Repeat substeps 4a through 4e, checking to the follow limits:
Rise time 35Ăns, maximum
Aberrations 0.35Ădiv., maximum