User's Manual
Performance Tests
Performance Verification
4Ć34
7. Disconnect the DMM.
This procedure checks the pulse response characteristics of the AWG2005
output waveforms at amplitudes of 0.5 and 1ĂV.
Electrical Characteristic Checked: Main Output, Pulse Response, on
pageĂ1Ć12.
Equipment Required: A 50ĂΩ coaxial cable and an oscilloscope.
Prerequisites: The AWG2005 meets the prerequisites listed on page 4Ć7.
Procedure:
1.
a. Connect the AWG2005 CH1 output
connector through the coaxial cable to the CH1 vertical input conĆ
nector on the oscilloscope (see Figure 4Ć15).
OscilloscopeAWG2005
AWG2005
Figure 4Ć15:ăPulse Response Initial Test Hookup
b.
Vertical CH1
Coupling DC
Scale 0.1ĂV/div.
Input impedance 50ĂΩ
Horizontal
Sweep 20Ăns/div.
Trigger
Source CH1
Coupling DC
Slope Positive
Level 0ĂV
Mode Auto
Pulse Response