User's Manual

Performance Tests
AWG2005 Service Manual
4Ć31
Procedure:
1. Install test hookup and set controls:
a. Hook up DMM: Connect the AWG2005 CH1 output through a 50Ă
coaxial cable, a 50 termination, and a dual banana connector to
the DMM INPUT connector (see Figure 4Ć13).
DMM
AWG2005
8842A
AWG2005
Figure 4Ć13:ăGain Accuracy Initial Test Hookup
b. Set DMM controls:
Mode VDC
Range 20
Input Front
2. Set the AWG2005 controls and select the waveform file:
a. Initialize AWG2005 controls: Push UTILITYMiscConfig...ReĆ
set to FactoryO.K.
b. Select the AWG2005 waveform file:
Push MODEAutostepSelect Autostep Fil.
Turn the general purpose knob to select the GAIN.AST file.
Push ENTER to select the file.
3. Check gain accuracy:
Check that the displayed step is Step 1 on the MODE menu. If it
is not, select the side menu STOP button to return to Step 1.
Note the DMM reading as "A" for this value.
Push the AWG2005 MANUAL TRIGGER button, and check that
the displayed step is Step 2.
Note the DMM reading as "B" for this value.
Do the following calculation.
A-B
Check that the calculated value is in the range from 0.990 V to
1.010 V.