Technical Reference TDS 500C, TDS 600B & TDS 700C Digitizing Oscilloscopes Performance Verification and Specifications 070-9874-02
Copyright Tektronix, Inc. All rights reserved. Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supercedes that in all previously published material. Specifications and price change privileges reserved. Printed in the U.S.A. Tektronix, Inc., P.O. Box 1000, Wilsonville, OR 97070–1000 TEKTRONIX and TEK are registered trademarks of Tektronix, Inc.
WARRANTY Tektronix warrants that this product will be free from defects in materials and workmanship for a period of three (3) years from the date of shipment. If any such product proves defective during this warranty period, Tektronix, at its option, either will repair the defective product without charge for parts and labor, or will provide a replacement in exchange for the defective product.
Table of Contents General Safety Summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . v Preface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ix Contacting Tektronix . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . x Performance Verification Procedures Performance Verification Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . 1–1 Input Channels vs.
Table of Contents List of Figures Figure 1–1: Map of display functions . . . . . . . . . . . . . . . . . . . . . . . . . . Figure 1–2: Verifying adjustments and signal-path compensation . . Figure 1–3: Universal test hookup for functional tests – TDS 600B shown . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Figure 1–4: Measurement of DC offset accuracy at zero setting . . . . Figure 1–5: Initial test hookup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Table of Contents Figure 1–33: Subsequent 60 Hz Rejection test hookup . . . . . . . . . . . . Figure 1–34: 60 Hz Rejection test result – TDS 684B shown . . . . . . . Figure 1–35: Line count accuracy test hookup . . . . . . . . . . . . . . . . . . . Figure 1–36: Line count accuracy test setup waveform – TDS 684B shown . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Figure 1–37: Line count accuracy correct result waveform . . . . . . . .
Table of Contents List of Tables Table 1–1: Test equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Table 1–2: DC offset accuracy (zero setting) . . . . . . . . . . . . . . . . . . . . Table 1–3: DC Voltage measurement accuracy . . . . . . . . . . . . . . . . . . Table 1–4: Analog bandwidth . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Table 1–5: Delay between channels worksheet . . . . . . . . . . . . . . . . . . Table 1–6: Available Filters . . . . . . . . .
General Safety Summary Review the following safety precautions to avoid injury and prevent damage to this product or any products connected to it. To avoid potential hazards, use the product only as specified. Only qualified personnel should perform service procedures. While using this product, you may need to access other parts of the system. Read the General Safety Summary in other system manuals for warnings and cautions related to operating the system.
General Safety Summary Product Damage Precautions Use Proper Power Source Do not operate this product from a power source that applies more than the voltage specified. Provide Proper Ventilation To prevent product overheating, provide proper ventilation. Do Not Operate With Suspected Failures Do Not Immerse in Liquids If you suspect there is damage to this product, have it inspected by qualified service personnel. Clean the probe using only a damp cloth. Refer to cleaning instructions.
General Safety Summary Symbols on the Product The following symbols may appear on the product: DANGER High Voltage Protective Ground (Earth) Terminal ATTENTION Refer to Manual Double Insulated Certifications and Compliances CSA Certified Power Cords CSA Certification includes the products and power cords appropriate for use in the North America power network. All other power cords supplied are approved for the country of use.
General Safety Summary viii TDS 500C, TDS 600B & TDS 700C Performance Verification and Specifications
Preface This is the Performance Verification and Specifications for the TDS 500C, TDS 600B, and TDS 700C Oscilloscopes. It contains procedures suitable for determining if each instrument functions, was adjusted properly, and meets the performance characteristics as warranted.
Preface Contacting Tektronix Product Support For application-oriented questions about a Tektronix measurement product, call toll free in North America: 1-800-TEK-WIDE (1-800-835-9433 ext. 2400) 6:00 a.m. – 5:00 p.m. Pacific time Or, contact us by e-mail: tm_app_supp@tek.com For product support outside of North America, contact your local Tektronix distributor or sales office. Service Support Contact your local Tektronix distributor or sales office.
Performance Verification Procedures
Performance Verification Procedures Two types of Performance Verification procedures can be performed on this product: Brief Procedures and Performance Tests. You may not need to perform all of these procedures, depending on what you want to accomplish. H To rapidly confirm that the oscilloscope functions and was adjusted properly, just do the brief procedures under Self Tests, which begin on page 1–5.
Performance Verification Procedures Input Channels vs. Model When performing the procedures in this section, be aware that some TDS models refer to input channels Ch 3 and Ch 4 as Aux 1 and Aux 2 respectively. Where appropriate, both names will appear in the procedure, for example, Ch 3 (Aux 1). The channel names for the various TDS models are shown below.
Performance Verification Procedures Initialize the oscilloscope: Press save/recall SETUP. Now, press the main-menu button Recall Factory Setup; then the side-menu button OK Confirm Factory Init. H Where instructed to use a front-panel button or knob, or select from a main or side menu, or verify a readout or status message, the name of the button or knob appears in boldface type: “press SHIFT; then UTILITY, press the main-menu button System until Cal is highlighted in the pop-up menu.
Performance Verification Procedures 1–4 TDS 500C, TDS 600B & TDS 700C Performance Verification and Specifications
Brief Procedures The Self Tests use internal routines to confirm basic functionality and proper adjustment. No test equipment is required to do these test procedures. The Functional Tests utilize the probe-compensation output at the front panel as a test-signal source for further verifying that the oscilloscope functions properly. A probe is required to do these test procedures. Self Tests This procedure uses internal routines to verify that the oscilloscope functions and was adjusted properly.
Brief Procedures f. H Press SHIFT; then press UTILITY. H Highlight Cal in the pop-up menu by repeatedly pressing the main-menu button System. See Figure 1–2. H Verify that the word Pass appears in the main menu under the following menu labels: Voltage Reference, Frequency Response, and Pulse Trigger. See Figure 1–2. Run the signal-path compensation: Press the main-menu button Signal Path; then press the side-menu button OK Compensate Signal Paths. g.
Brief Procedures Functional Tests The purpose of these procedures is to confirm that the oscilloscope functions properly. The only equipment required is one of the standard-accessory probes and, to check the file system, a 3.5 inch, 720 K or 1.44 Mbyte floppy disk. CAUTION. The P6243 and P6245 probes that can be used with this oscilloscope provide an extremely low loading capacitance (<1 pF) to ensure the best possible signal reproduction.
Brief Procedures connect the probe ground (typically black) to PROBE COMPENSATION GND. If using a P6243 or P6245 probe, you may want to attach a Y-lead connector and two SMD KlipChips as shown in Figure 1–3. Signal Gnd Figure 1–3: Universal test hookup for functional tests – TDS 600B shown b. Initialize the oscilloscope: H Press save/recall SETUP. H Press the main-menu button Recall Factory Setup. H Press the side-menu button OK Confirm Factory Init. 2.
Brief Procedures H Move the probe to the channel you selected. b. Match the trigger source to the channel selected: H Press TRIGGER MENU. H Press the main-menu button Source. H Press the side-menu button that corresponds to the channel selected, Ch2, Ch3, or Ch4. (Some TDS models use Ax1 and Ax2 instead of Ch3 and Ch4). c. Set up the selected channel: H Set the vertical SCALE to 200 mV. H Set the horizontal SCALE to 200 s. Press CLEAR MENU to remove any menu that may be on the screen.
Brief Procedures f. H Envelope mode displays an actively acquiring waveform on-screen with the noise displayed. H Average mode displays an actively acquiring waveform on-screen with the noise reduced. Test all channels: Repeat substeps a through e until all four input channels are verified. 3. Remove the test hookup: Disconnect the probe from the channel input and the probe-compensation terminals.
Brief Procedures c. The horizontal POSITION knob positions the signal left and right on-screen when rotated. 3. Remove the test hookup: Disconnect the probe from the channel input and the probe-compensation terminals. Verify the Main and Delayed Trigger Systems Equipment required One probe such as the P6243, P6245, or P6139A Prerequisites None 1. Install the test hookup and preset the oscilloscope controls: a. Hook up the signal source: Install the probe on CH 1.
Brief Procedures H Pressing SET LEVEL TO 50% triggers the signal that you just left untriggered. (Leave the signal triggered). 3. Verify that the delayed trigger system operates: a. Select the delayed time base: H Press HORIZONTAL MENU. H Press the main-menu button Time Base. H Press the side-menu button Delayed Triggerable; then press the side-menu button Delayed Only. H Set the horizontal SCALE for the D (delayed) time base to 200 s. b.
Brief Procedures Verify the File System Equipment required One probe such as the P6243, P6245 or P6139A Prerequisites None One 720 K or 1.44 Mbyte, 3.5 inch DOS-compatible disk. You can use a disk of your own or you can use the Programming Examples Software 3.5 inch disk (Tektronix part number 063-1134-XX) contained in the TDS Family Programmer Manual (Tektronix part number 070-9556-XX). 1. Install the test hookup and preset the oscilloscope controls: a.
Brief Procedures H f. Turn the general purpose knob to select the file to save. Choose (or fd0:, the floppy disk drive, or hd0:, the optional hard disk drive). With this choice, you will save a file starting with , then containing 5-numbers, and a extension. For example, the first time you run this on a blank, formatted disk or on the Example Programs Disk, the oscilloscope will assign the name to your file.
Performance Tests This section contains a collection of procedures for checking that the TDS 500C, TDS 600B, and TDS 700C Digitizing Oscilloscopes perform as warranted. The procedures are arranged in four logical groupings: Signal Acquisition System Checks, Time Base System Checks, Triggering System Checks, and Output Ports Checks. They check all the characteristics that are designated as checked in Specifications.
Performance Tests Equipment Required These procedures use external, traceable signal sources to directly check warranted characteristics. The required equipment list follows this introduction. Table 1–1: Test equipment Item number and description Minimum requirements Example Purpose 1. Attenuator,10X (two required) Ratio: 10X; impedance 50 ; connectors: female BNC input, male BNC output Tektronix part number 011-0059-02 Signal Attenuation 2.
Performance Tests Table 1–1: Test equipment (cont.) Item number and description Minimum requirements Example Purpose 14. Generator, Video Signal Provides PAL compatible outputs Tektronix TSG 121 Used to Test Video Option 05 Equipped Instruments Only 15.
Performance Tests Table 1–1: Test equipment (cont.) Item number and description Minimum requirements Example Purpose 27. Generator, Optical Impulse 850 nm optical impulse Tektronix OIG501 Optical Impulse Generator1 Checking Option 3C 28. Generator, Optical Impulse 1300 nm optical impulse Tektronix OIG502 Optical Impulse Generator1 Checking Option 4C 29. Attenuator, Optical 62.5 micron optical attenuator Tektronix OA5022 Optical Attenuator1 Checking Option 3C and 4C 30.
Performance Tests TDS 600B Test Record Photocopy this and the next three pages and use them to record the performance test results for your TDS 600B. TDS 600B Test Record Instrument Serial Number: Temperature: Date of Calibration: TDS 600B Performance Test Certificate Number: RH %: Technician: Minimum Incoming Outgoing Maximum Offset Accuracy CH1 Offset +1 mV +101 mV +1.01 V – 2.1 mV – 75.6 mV – 756 mV __________ __________ __________ __________ __________ __________ + 2.1 mV + 75.
Performance Tests TDS 600B Test Record (Cont.) Instrument Serial Number: Temperature: Date of Calibration: Certificate Number: RH %: Technician: TDS 600B Performance Test Minimum Incoming Outgoing Maximum CH2 200 mV Vert scale setting, +5 Div position setting, –10 V offset – 11.7835 V __________ __________ – 11.4165 V CH2 1 V Vert scale setting, –5 Div position setting, +10 V offset + 17.2425 V __________ __________ + 18.
Performance Tests TDS 600B Test Record (Cont.) Instrument Serial Number: Temperature: Date of Calibration: TDS 600B Performance Test Certificate Number: RH %: Technician: Minimum Incoming Outgoing Maximum N/A __________ __________ 100 ps –2.0 Div __________ __________ +2.0 Div Pulse-Glitch or Pulse-Width, Hor. scale ≤ 1 ms Lower Limit Upper Limit 3.5 ns 3.5 ns __________ __________ __________ __________ 6.5 ns 6.5 ns Pulse-Glitch or Pulse-Width, Hor.
Performance Tests TDS 600B Test Record (Cont.) Instrument Serial Number: Temperature: Date of Calibration: Certificate Number: RH %: Technician: TDS 600B Performance Test Minimum Incoming Outgoing Maximum CH 3 or AX1 SIGNAL OUTPUT, 1 M Pk-Pk ≥ 80 mV __________ __________ Pk-Pk ≤ 120 mV CH 3 or AX1 SIGNAL OUTPUT, 50 Pk-Pk ≥ 40 mV __________ __________ Pk-Pk ≤ 60 mV Frequency (CH1 Freq). 950 Hz __________ __________ 1.
Performance Tests TDS 500C/700C Test Record Photocopy this and the next three pages and use them to record the performance test results for your TDS 500C/700C. TDS 500C/700C Test Record Instrument Serial Number: Temperature: Date of Calibration: TDS 500C/700C Performance Test Certificate Number: RH %: Technician: Minimum Incoming Outgoing Maximum Offset Accuracy CH1 Offset +1 mV +101 mV +1.01 V – 1.6 mV – 25.1 mV – 251 mV __________ __________ __________ __________ __________ __________ + 1.
Performance Tests TDS 500C/700C Test Record (Cont.) Instrument Serial Number: Temperature: Date of Calibration: Certificate Number: RH %: Technician: TDS 500C/700C Performance Test Minimum Incoming Outgoing Maximum CH2 200 mV Vert scale setting, +5 Div position setting, –10 V offset – 11.6805 V __________ __________ – 11.5195 V CH2 1 V Vert scale setting, –5 Div position setting, +10 V offset + 17.7575 V __________ __________ + 18.
Performance Tests TDS 500C/700C Test Record (Cont.) Instrument Serial Number: Temperature: Date of Calibration: TDS 500C/700C Performance Test Certificate Number: RH %: Technician: Minimum Incoming Outgoing Maximum N/A __________ __________ 50 ps –2.5 Div __________ __________ +2.5 Div Pulse-Glitch or Pulse-Width, Hor. scale ≤ 1 ms Lower Limit Upper Limit 3.5 ns 3.5 ns __________ __________ __________ __________ 6.5 ns 6.5 ns Pulse-Glitch or Pulse-Width, Hor.
Performance Tests TDS 500C/700C Test Record (Cont.) Instrument Serial Number: Temperature: Date of Calibration: Certificate Number: RH %: Technician: TDS 500C/700C Performance Test Minimum Incoming Outgoing Maximum CH 3 or AX 1 SIGNAL OUTPUT, 1 M Pk-Pk ≥ 88 mV __________ __________ Pk-Pk ≤ 132 mV CH 3 or AX 1 SIGNAL OUTPUT, 50 Pk-Pk ≥ 44 mV __________ __________ Pk-Pk ≤ 66 mV Frequency (CH1 Freq). 950 Hz __________ __________ 1.
Performance Tests Signal Acquisition System Checks These procedures check those characteristics that relate to the signal-acquisition system and are listed as checked under Warranted Characteristics in Specifications. Check Offset Accuracy (Zero Setting) Equipment required None Prerequisites The oscilloscope must meet the prerequisites listed on page 1–15. See Input Channels vs. Model on page 1–2. 1. Preset the instrument controls: a. Initialize the oscilloscope: H Press save/recall SETUP.
Performance Tests b. Set the vertical scale: Set the vertical SCALE to one of the settings listed in Table 1–2 that is not yet checked. (Start with the first setting listed). H Press VERTICAL MENU. Press the main-menu button Fine Scale. H Use the keypad to enter the vertical scale. For the 1 mV setting, press 1, SHIFT, m, then ENTER. For the 101 mV setting, press 101, SHIFT, m, then ENTER. For the 1.01 V setting, press 1.01, then ENTER. H Press CLEAR MENU.
Performance Tests Read the measurement results. 2 Align the active cursor over the waveform. 1 Figure 1–4: Measurement of DC offset accuracy at zero setting g. Test all channels: Repeat substeps a through f for all input channels. 3. Disconnect the hookup: No hookup was required. Check DC Voltage Measurement Accuracy WARNING. The generator is capable of outputting dangerous voltages.
Performance Tests 1. Install the test hookup and preset the instrument controls: a. Hook up the test-signal source: H Set the output of a DC calibration generator to 0 volts. H Connect the output of a DC calibration generator through a dual-banana connector followed by a 50 W precision coaxial cable to one side of a BNC T connector. See Figure 1–5.
Performance Tests a. Select an unchecked channel: H Press WAVEFORM OFF to remove the channel just confirmed from the display. H Press the front-panel button that corresponds to the channel you are to confirm. H Set the generator output to 0 V. H Move the test hookup to the channel you selected. b. Turn on the measurement Mean for the channel: H Press MEASURE, then press the main-menu button Select Measrmnt for CHx.
Performance Tests H Use the keypad to set vertical offset to the positive-polarity setting listed in the table for the current vertical scale setting. The baseline level will remain off screen. H Set the generator to the level and polarity indicated in the table for the vertical scale, position, and offset settings you have made. The DC test level should appear on screen. (If it doesn’t return, the DC accuracy check is failed for the current vertical scale setting of the current channel). e.
Performance Tests H Repeat substeps c through f until all vertical scale settings, listed in Table 1–3, are checked for the channel under test. g. Test all channels: Repeat substeps a through f for all four channels. 3. Disconnect the hookup: a. Set the generator output to 0 V. b. Disconnect the cable from the generator output at the input connector of the channel last tested.
Performance Tests NOTE. Refer to the Sine Wave Generator Leveling Procedure on page 1–84 if your sine wave generator does not have automatic output amplitude leveling. c. Hook up the test-signal source: Connect the sine wave output of a leveled sine wave generator to CH 1. Set the output of the generator to a reference frequency of 10 MHz or less. See Figure 1–7.
Performance Tests H Press the front-panel button that corresponds to the channel you are to confirm. H Move the leveling output of the sine wave generator to the channel you selected. b. Match the trigger source to the channel selected: H Press TRIGGER MENU. Press the main-menu button Source; then press the side-menu button that corresponds to the channel selected. c. Set its input impedance: H Press VERTICAL MENU; then press the main-menu button Coupling.
Performance Tests f. 1–36 H Press the side-menu button more, if needed, until the menu label Frequency appears in the side menu (its icon is shown at the left). Press the side-menu button Frequency. H Press the side-menu button more until the menu label Pk-Pk appears in the side menu (its icon is shown at the left). Press the side-menu button Pk-Pk. H Press CLEAR MENU.
Performance Tests Read results. 3 Set the generator (reference) frequency to the test frequency from Table 1–4. 1 Set the horizontal scale from Table 1–4. 2 Figure 1–9: Measurement of analog bandwidth g. Check against limits: H CHECK that the Pk-Pk readout on screen is within the limits listed in Table 1–4 for the current vertical scale setting. H Enter voltage on test record. H When finished checking, set the horizontal SCALE back to the 50 ns setting. STOP.
Performance Tests NOTE. Passing the signal path compensation confirms the signal path for all vertical scale settings for all channels. Passing the internal diagnostics ensures that the factory-set adjustment constants that control the bandwidth for each vertical scale setting have not changed. h. Check remaining vertical scale settings against limits (optional): i.
Performance Tests b. Modify the initialized front-panel control settings: H Do not adjust the vertical position of any channel during this procedure. H Set the horizontal SCALE to 500 ps. H Press SHIFT; then press ACQUIRE MENU. H Press the main-menu button Mode, and then press the side-menu button Average 16. c.
Performance Tests b. The horizontal SCALE should already be set to 500 ps. On the TDS 784C, and 600B, now set it to 200 ps. On the TDS 520C, 540C, 724C, and 754C, push the front-panel ZOOM button, press the side-menu On button, set the horizontal SCALE to 250 ps, and be sure the vertical scale factor is kept at 1.0X and the horizontal scale factor is 2.0X. c. Save a CH 2 waveform: Press CH 2. Be sure the vertical scale factor is kept at 1.0X. Then press save/recall WAVEFORM.
Performance Tests Read results. 4 Display the waveforms. 1 H Align one V bar cursor to the time reference point of the left-most waveform edge and the other cursor to the time reference point of the right-most waveform edge by rotating the General Purpose knob. (Press SELECT to switch between the two cursors). See Figure 1–11 on page 1–41. H Read the measurement results at the D: cursor readout, not the @: readout on screen. 2 3 Locate the time reference points for these waveforms.
Performance Tests j. Repeat the procedure from step 1.c through 2.e. k. Again use the cursors to measure the skew from CH1 to CH2, CH1 to CH3, and CH1 to CH4. Write down these numbers in the second measurement column of Table 1–5. Note that these numbers may be either positive or negative. l. Add the first CH1 to CH2 skew measurement to the second CH1 to CH2 skew measurement and divide the result by 2. Use Table 1–5. m.
Performance Tests Time Base System Checks These procedures check those characteristics that relate to the Main and Delayed time base system and are listed as checked under Warranted Characteristics in Specifications.
Performance Tests H Use the vertical POSITION knob to center the test signal on screen. H Set the horizontal SCALE of the Main time base to 1 ms. H Press TRIGGER MENU; then press the main-menu button Mode & Holdoff. Press the side-menu button Normal. 2. Confirm Main and Delayed time bases are within limits for accuracies: a. Display the test signal: H Align the trigger T to the center vertical graticule line by adjusting the horizontal POSITION. See Figure 1–13 on page 1–45.
Performance Tests Align the trigger T to the center graticule line. 1 Set 2 horizontal mode. 4 3 Check long–term sample rates and delay time accuracies against limits. Set horizontal scale and delayed time. Figure 1–13: Measurement of accuracy — Long-term and delay time 3. Disconnect the hookup: Disconnect the cable from the generator output at the input connector of CH 1.
Performance Tests 1. Install the test hookup and preset the instrument controls: a. Initialize the instrument: H Press save/recall SETUP. H Press the main-menu button Recall Factory Setup. H Press the side-menu button OK Confirm Factory Init. b. Modify the default setup: H Press VERTICAL MENU. H Press the main-menu button Coupling; then press the side-menu button to select 50 coupling. H Set the horizontal SCALE to 10 ns on the TDS 600B and 12.5 ns on the TDS 500C/700C. c.
Performance Tests b. Set the trigger mode: Press TRIGGER MENU. Now press the main-menu button Mode & Holdoff; then press the side-menu button Normal. c. Set upper and lower limits that ensure triggering: See Figure 1–15. H Press the main-menu button Type; then repeatedly press the same button until Pulse is highlighted in the menu that pops up. H Press the main-menu button Class; then repeatedly press the same button until Width is highlighted in the menu that pops up.
Performance Tests Set upper and lower limits that ensure triggering. Then change limits until triggering stops. Figure 1–15: Measurement of time accuracy for pulse and glitch triggering 3. Confirm the trigger system is within time-accuracy limits for pulse-glitch or pulse-width triggering (horizontal scale >1 s): a. Set upper and lower limits that ensure triggering at 250 kHz: H Press the side-menu button Upper Limit. Use the keyboard to set the upper limit to 4 s.
Performance Tests H CHECK that the Lower Limit readout, after the oscilloscope stops triggering, is within 1.9 s to 2.1 s, inclusive. H Enter time on test record. H Use the keypad to return the Lower Limit to 500 ns and reestablish triggering. H Press the side-menu button Upper Limit; then use the general purpose knob to slowly decrease the Upper Limit readout until triggering stops. H CHECK that the Upper Limit readout, after the oscilloscope loses triggering, is within 1.9 s to 2.
Performance Tests H Connect the output of the DC calibration generator, through a dual-banana connector followed by a 50 W precision coaxial cable, to one side of a BNC T connector. H Connect the Sense output of the generator, through a second dual-banana connector followed by a 50 W precision coaxial cable, to other side of the BNC T connector. Now connect the BNC T connector to CH 1. b. Initialize the oscilloscope: H Press save/recall Setup. H Press the main-menu button Recall Factory Setup.
Performance Tests c. Read results (Check against limits): See Figure 1–17. Set vertical position to –3 divs. Set vertical offset to +10 volts. H CHECK that the Level readout in the main menu is within 9.863 V to 10.137 V, inclusive, for the TDS 600B or is within 9.9393 V to 10.1147 V, inclusive, for the TDS 500C/700C. H Enter voltage on test record. H Press the main-menu button Slope; then press the side-menu button for negative slope. See icon at left. Repeat substep b.
Performance Tests H Press HORIZONTAL MENU. H Press the main-menu button Time Base. H Press the side-menu buttons Delayed Only and Delayed Triggerable. H Set D (delayed) horizontal SCALE to 500 s. b. Select the Delayed trigger system: H Press SHIFT; then press the front-panel DELAYED TRIG button. H Press the main-menu button Level. c. Measure the test signal: Press the side-menu button SET TO 50%. Read the measurement results in the side (or main) menu below the label Level. d.
Performance Tests Sensitivity, Edge Trigger, DC Coupled Equipment required One sine wave generator (Item 19 or, optionally, items 25 and 26) Two precision 50 coaxial cables (Item 5) One 10X attenuator (Item 1) One BNC T connector (Item 7) One 5X attenuator (Item 2) Prerequisites See page 1–15. 1. Install the test hookup and preset the instrument controls: a. Initialize the oscilloscope: H Press save/recall SETUP. H Press the main-menu button Recall Factory Setup.
Performance Tests Sine Wave Generator To AUX TRIG INPUT on rear panel Digitizing Oscilloscope Figure 1–18: Initial test hookup 2. Confirm Main and Delayed trigger systems are within sensitivity limits (50 MHz): a. Display the test signal: H Set the generator frequency to 50 MHz. H Press MEASURE. H Press the main-menu button Level Setup; then press the side-menu button Min-Max. H Press the main-menu button Select Measrmnt for Ch1.
Performance Tests 2 ms/division and faster, TRIG’D will remain constantly lighted. It will flash for slower settings. H Press TRIGGER MENU; then press the main-menu button Slope. H Press SET LEVEL TO 50%. Adjust the TRIGGER LEVEL knob so that the TRIG’D light is on. Set the level to near the middle of the range where the TRIG’D light is on. CHECK that the trigger is stable for the test waveform on both the positive and negative slopes. Use the side menu to switch between trigger slopes.
Performance Tests H Press SHIFT; then press DELAYED TRIG. Press the main-menu button Level. H Press the side-menu button SET TO 50%. CHECK that a stable trigger is obtained for the test waveform for both the positive and negative slopes of the waveform. Use the TRIGGER LEVEL knob to stabilize the Main trigger. Use the general purpose knob to stabilize the Delayed trigger. Press the main-menu button Slope; then use the side menu to switch between trigger slopes. See Figure 1–19.
Performance Tests H Leave the Main trigger system triggered on the positive slope of the waveform before proceeding to the next check. H Press the main-menu button Source; then press the side-menu button –more– until CH 1 appears. Press CH 1. 4. Confirm that the Main and Delayed trigger systems are within sensitivity limits (full bandwidth): a. Hook up the test-signal source: Disconnect the hookup installed in step 1.
Performance Tests Output Signal Checks The procedure that follows checks those characteristics of the output signals that are listed as checked under Warranted Characteristics in Specifications. The oscilloscope outputs these signals at its front and rear panels. Check Outputs — CH 3 (AX1 on some models) Main and Delayed Trigger Equipment required Two 50 precision cables (Item 5) Prerequisites See page 1–15.
Performance Tests d. Modify the initialized front-panel control settings: H Set the horizontal SCALE to 200 ms. H Press SHIFT; then press ACQUIRE MENU. H Press the main-menu button Mode; then press the side-menu button Average. H Select 64 averages. Do this with the keypad or the general purpose knob 2. Confirm Main and Delayed Trigger outputs are within limits for logic levels: a. Display the test signal: H Press WAVEFORM OFF to turn off CH 1. H Press CH 2 to display that channel.
Performance Tests Check output Figure 1–21: Measurement of main trigger out limits d. Check Delayed Trigger output against limits: See Figure 1–21. H Move the precision 50 W cable from the rear-panel Main Trigger Output BNC to the rear-panel Delayed Trigger Output BNC. H CHECK that the Ch2 High readout is ≥1.0 volt and that the Ch2 Low readout ≤0.25 volts. H Enter high and low voltages on test record. H Press the side-menu button to select the 1 MW setting. H Press CLEAR MENU.
Performance Tests H Press the main-menu button Source. H Press the side-menu button Ch3. (Ax1 on some TDS models) H Set vertical SCALE to 100 mV. H Press SET LEVEL TO 50%. H Press MEASURE; then press the main-menu button Select Measrmnt for Ch2. H Repeatedly press the side-menu button –more– until Pk-Pk appears in the side menu (its icon is shown at the left). Press the side-menu button Pk-Pk. H Press CLEAR MENU. b.
Performance Tests Check Probe Compensator Output Equipment required One female BNC to clip adapter (Item 3) Two dual-banana connectors (Item 6) One BNC T connector (Item 7) Two 50 precision cables (Item 5) One DC calibration generator (Item 9) Prerequisites See page 1–15. Also, the digitizing oscilloscope must have passed Check Accuracy For Long-Term Sample Rate, Delay Time, and Delta Time Measurements on page 1–43. 1. Install the test hookup and preset the instrument controls: a.
Performance Tests c. Modify the initialized front-panel control settings: H Set the horizontal SCALE to 200 s. H Press SET LEVEL TO 50%. H Use the vertical POSITION knob to center the display on screen. H Press SHIFT; then press ACQUIRE MENU. H Press the main-menu button Mode; then press the side-menu button Average. H Select 128 averages with the keypad or the general purpose knob. 2. Confirm that the Probe Compensator signal is within limits for frequency: a.
Performance Tests Figure 1–23: Measurement of probe compensator frequency c. Save the probe compensation signal in reference memory: H Press SAVE/RECALL WAVEFORM; then press the main-menu button Save Wfm Ch 1. H Press the side-menu button To Ref 1 to save the probe compensation signal in reference 1. H Disconnect the cable from CH 1 and the clips from the probe compensation terminals. H Press MORE; then press the main-menu button Ref 1 to displayed the stored signal. H Press CH 1. d.
Performance Tests H Connect the Sense output of the generator through a second dual-banana connector followed by a 50 W precision coaxial cable to the other side of the BNC T connector. Now connect the BNC T connector to CH 1. See Figure 1–24. Digitizing Oscilloscope DC Calibrator Dual banana to BNC adapters 50 coaxial cables BNC T connector Figure 1–24: Subsequent test hookup e. Measure amplitude of the probe compensation signal: f. H Press SHIFT; then press ACQUIRE MENU.
Performance Tests Figure 1–25: Measurement of probe compensator amplitude g. Check against limits: H Subtract the value just obtained (base level) from that obtained previously (top level). H CHECK that the difference obtained is within 495 mV to 505 mV, inclusive. H Enter voltage difference on test record. 3. Disconnect the hookup: Disconnect the cable from CH 1.
Performance Tests Option 05 Video Trigger Checks Check Video Trigger Equipment required PAL signal source (Item 14) 60 Hz. sine wave generator (Item 15) Pulse generator (Item 16) Two 75 cables (Item 17) Two 75 terminators (Item 18) One BNC T connector (Item 7) 50 cable (Item 5) 50 terminator (Item 4) Prerequisites See page 1–15. These prerequisites include running the signal path compensation routine. 1. Set up digitizing oscilloscope to factory defaults by completing the following steps: a.
Performance Tests g. Press the main-menu Bandwidth. h. Select 250 MHz from the side menu. i. Press the main-menu Fine Scale. j. Use the keypad to set the fine scale to 282mV (press 282, SHIFT, m, then ENTER). k. Press HORIZONTAL MENU. l. Press the main-menu Horiz Scale. m. Use the keypad to set the horizontal scale to 200 ns (press 200, SHIFT, n, then ENTER). 3. Check Jitter vs. Signal Amplitude a. Set up equipment for Jitter Test. See Figure 1–26.
Performance Tests Figure 1–27: Jitter test displayed waveform – TDS 684B shown c. Press SHIFT; then press ACQUIRE MENU. d. Press the main-menu Mode. e. Select the side-menu Average. It should be already set to 16. f. Press the main-menu Create Limit Test Template. g. Press the side-menu V Limit. h. Use the keypad to set V Limit to 180 mdiv (press 180, SHIFT, m, then ENTER). i. Press the side-menu OK Store Template. j. Press MORE. k. Press the main-menu Ref1. l. Press CH1. m.
Performance Tests q. Press the main-menu Mode. r. Press the side-menu Envelope. s. Use the keypad to set envelope to use 100 acquisitions (press 100, then ENTER). t. Press the main-menu Stop After button. u. Press the side-menu Single Acquisition Sequence. v. Confirm that the oscilloscope successfully makes 100 acquisitions. If not successful, the oscilloscope bell will ring. When the word Run in the top left corner of the display changes to STOP, the test is complete. See Figure 1–28.
Performance Tests b. Press WAVEFORM OFF. c. Press HORIZONTAL MENU. d. Use the keypad to set horizontal scale (/div) to 50 s (press 50, SHIFT, m, then ENTER). e. Press SHIFT; then press ACQUIRE MENU. f. Press the main-menu Stop After. g. Press the side-menu RUN/STOP button only. h. Press the main-menu Mode. i. Press the side-menu Sample. j. Press RUN/STOP. k. Press VERTICAL MENU. l. Use the keypad to set fine scale to 300 mV (press 300, SHIFT, m, then ENTER). m.
Performance Tests Positive pulse Negative pulses Figure 1–29: Triggered signal range test – 300 mV n. Use the keypad to set the fine scale to 75 mV (press 75, SHIFT, m, then ENTER). o. CONFIRM that the TRIG’D LED stays lighted and that the waveform on screen is stable. That is, it does not move horizontally or vertically. Also, CONFIRM that the waveform on the screen has one positive pulse and a number of negative pulses. See Figure 1–30.
Performance Tests Positive pulse Negative pulses Figure 1–30: Triggered signal range test – 75 mV p. Disconnect all test equipment (TSG121) from the digitizing oscilloscope. 5. Check 60 Hz Rejection. a. Set up oscilloscope for 60 Hz Rejection Test: H Use the keypad to set the Ch1 Fine Scale to 282 mV (press 282, SHIFT m, then ENTER). H Press WAVEFORM OFF. H Press CH2. H Press VERTICAL MENU. H Use the keypad set the fine scale to 2 V (press 2, then ENTER). H Press HORIZONTAL MENU.
Performance Tests Digitizing Oscilloscope Signal Generator Output 50 cable Figure 1–31: 60 Hz Rejection test hookup H Adjust the signal generator for three vertical divisions of 60 Hz signal. See Figure 1–32. The signal will not be triggered. That is, it will run free. Figure 1–32: 60 Hz Rejection test setup signal c. Check 60 Hz rejection: H 1–74 Use the keypad to set the horizontal scale (/div) to 50 s (press 50, SHIFT, m, then ENTER).
Performance Tests H Reconnect the output of the signal generator. Connect the composite signal connector of the PAL signal source (labeled COMPST on the TSG 121) to a 75 W cable and a 75 W terminator. Connect both signals to the CH1 input through a BNC T. See Figure 1–33. H Press VERTICAL MENU. H If needed, press the main-menu Fine Scale. H Use the keypad to set fine scale to 500 mV (press 500, SHIFT, m, then ENTER).
Performance Tests Figure 1–34: 60 Hz Rejection test result – TDS 684B shown 6. Check Line Count Accuracy. a. Set up oscilloscope for Line Count Accuracy Test: H Press WAVEFORM OFF. H Press CH1. H Press HORIZONTAL MENU. H Press the main-menu Record Length. H Press the side-menu –more– until you see the appropriate menu. H Press the side-menu 5000 points in 100divs. H Press the main-menu Horiz Scale (/div).
Performance Tests PAL signal source Digitizing Oscilloscope Output 75 cable 75 terminator Figure 1–35: Line count accuracy test hookup H Press the main-menu Trigger Position. H Press the side-menu to Set to 50%. H Press the main-menu to Horiz Pos. H Press the side-menu to Set to 50%. H Use the HORIZONTAL POSITION knob to move the falling edge of the sync pulse to two divisions to the left of center screen. See Figure 1–36.
Performance Tests Figure 1–36: Line count accuracy test setup waveform – TDS 684B shown 1–78 H Press CURSOR. H Press the main-menu Function. H Press the side-menu V Bars. H Using the General Purpose knob, place the left cursor directly over the trigger ‘T’ icon. H Press SELECT. H Turn the General Purpose knob to adjust the right cursor for a cursor delta reading of 6.780us. H Use the HORIZONTAL POSITION knob to position the right cursor to center screen.
Performance Tests Figure 1–37: Line count accuracy correct result waveform H Disconnect all test equipment (TSG 121) from the digitizing oscilloscope. H Turn off cursors by pressing CURSOR, then the main-menu Function button, and, finally, Off from the side menu. 7. Check the Sync Duty Cycle. a. Set up digitizing oscilloscope for Sync Duty Cycle Test: H Press TRIGGER MENU. H Press the Standard pop-up to select FlexFmt. Trigger Type should already be set to Video. H Press the main-menu Setup.
Performance Tests H Press the side-menu Fields. H Use the keypad to set the number of fields to 1 (press 1, then ENTER). H Press the side-menu Sync Width. H Use the keypad to set the width to 400 ns (press 400, SHIFT, n, then ENTER). H Press the side-menu –more– 1 of 2. Then press V1 Start Time. H Use the keypad to set V1 start time to 10.10 s (press 10.10, SHIFT, m, then ENTER). H Press the side-menu V1 Stop Time. H Use the keypad to set V1 stop time to 10.50 s (press 10.
Performance Tests c. Check Sync Duty Cycle: H Connect the pulse generator through a 50 W cable and a 50 W terminator to the oscilloscope CH1 input. See Figure 1–38. Pulse Generator Digitizing Oscilloscope Output 50 cable 50 terminator Figure 1–38: Setup for sync duty cycle test H Turn the pulse generator OUTPUT (VOLTS) control until the signal on the oscilloscope shows a one division negative going pulse. See Figure 1–39. NOTE.
Performance Tests Figure 1–39: Sync duty cycle test: one-div neg pulse waveform H Turn the pulse generator PULSE DURATION variable control to adjust the negative pulse so the oscilloscope’s CH1 – Width measurement displays 400ns +/–10 ns. H Turn the HORIZONTAL SCALE knob to set the oscilloscope time base to 5ms/div. H Turn the pulse generator PERIOD variable control to adjust the period until the oscilloscope CH1 Period measurement reads 21.000ms –25/+50 ns. See Figure 1–40. Read note shown below.
Performance Tests Figure 1–40: Sync duty cycle test: critically adjusted pulse H Press TRIGGER MENU. H Press the main-menu Type pop-up until you select Video. If the TRIG’D LED is not lighted, check that the CH1 – Width and CH1 Period measurements are adjusted correctly. See note above. CONFIRM that the setup is correct and the oscilloscope will trigger. H CONFIRM that the TRIG’D LED is lighted and the waveform is stable.
Performance Tests H Disconnect the signal source from CH1, wait a few seconds, then reconnect the signal. H CONFIRM that the TRIG’D LED is lighted and the waveform is stable. H Disconnect all test equipment from the digitizing oscilloscope. H Press save/recall SETUP, the main-menu button Recall Factory Setup, and the side-menu OK Confirm Factory Init. Sine Wave Generator Leveling Procedure Some procedures in this manual require a sine wave generator to produce the necessary test signals.
Performance Tests Sine Wave Generator Digitizing Oscilloscope Level Meter Input Power splitter Output Attenuators (if necessary) Power sensor Figure 1–41: Sine wave generator leveling equipment setup 1. Install the test hookup: Connect the equipment as shown in Figure 1–41. 2. Set the Generator: H Set the sine wave generator to a reference frequency of 10 MHz. H Adjust the sine wave generator amplitude to the required number of divisions as measured by the digitizing oscilloscope. 3.
Performance Tests Equipment required Sine wave generator (Item 19) Level meter and power sensor (Item 20) Two male N to female BNC adapters (Item 23) Two precision coaxial cables (Item 5) Prerequisites See page 1–15 1. Install the test hookup: Connect the equipment as shown in Figure 1–42 (start with the sine wave generator connected to the digitizing oscilloscope).
Performance Tests 4. Set the generator to the new frequency and reference level: H Change the sine wave generator to the desired new frequency. H Input the correction factor for the new frequency into the level meter. H Adjust the sine wave generator amplitude until the level meter again reads the value noted in step 3. The signal amplitude is now correctly set for the new frequency. H Disconnect the sine wave generator from the power sensor.
Performance Tests Table 1–7: Reference Receiver Filter Options: Option 4C – SONET (P6703B 1300nm) Option 3C – Fibre Channel (P6701B 850nm) 52Mbit OC1 FC133Mbit FC266Mbit FC531Mbit 155Mbit OC3 155Mbit OC3 622Mbit OC12 622Mbit OC12 FC1063 (TDS784C Only) FC1063 (TDS784C Only) Table 1–8: Option 3C and 4C Specifications Name Description Calibration Range, Reference Receiver 10 mW per division 20 mW per division 50 mW per division Temperature Range, Warranted Filter 23_ C, "5_ C Calibration Range
Performance Tests Reference Receiver Verification Equipment required OIG501 (use with P6701B) (Item 27) OIG502 (use with P6703B) (Item 28) Optical Attenuator OA5022 (Item 29) Fibre Optic Cable (Item 30) Optical-to-Electrical Converter (item 31) Prerequisites See page 1–15. Also, the probe and the oscilloscope channel it is attached to must have been calibrated as a reference receiver. 1.
Performance Tests H Connect the optical probe on CH 1 of the oscilloscope to the OPTICAL OUTPUT of the optical attenuator using the fiber-optic cable of the probe. Set an optical impulse level that does not clip the vertical channel of the oscilloscope. c. Set up the OIG to run with LOW impulse energy. H Keep the impulse below 100 W peak level because high-energy impulses into the P670x probe will cause major changes in the frequency response of the probe. H Set the OIG for 1 Mhz internal trigger.
Performance Tests H Press HORIZONTAL MENU; then press the main-menu button Trigger Position. Set the Trigger Position to 20% using the general purpose knob or keypad. NOTE. Press SHIFT to change the horizontal position speed. The position moves faster when SHIFT is lighted. H Adjust the HORIZONTAL POSITION to move the 20% trigger point onto the display (see Figure 1–44). Figure 1–44: Optical impulse of Ch1 input from OA5022 Optical Attenuator (OIG501 / OIG502 fed into optical attenuator in Step 1) j.
Performance Tests NOTE. is the standard that you are verifying (such as OC1, OC3, OC12, FC1063; see Table 1–10). H From the Mask Type pop-up menu, select the mask type () of your reference receiver. H Press the Standard Mask main menu; then select your reference receiver mask () from the side menu. NOTE. For the Bessel Thompson (BT) filter to be active, the VERTICAL MENU deskew must be set to zero on all four channels. Also, the ACQUIRE MENU Acquisition Mode must be Sample.
Performance Tests 4. Move the beginning of the FFT data record onto the screen: NOTE. Press SHIFT to change the horizontal position speed. The position moves faster when SHIFT is lighted. H Adjust the HORIZONTAL POSITION control to move the beginning of the FFT data record onto the display. NOTE. Expanding the horizontal scale of the FFT waveform gives greater resolution when making cursor measurements.
Performance Tests Figure 1–45: Optical impulse response for OC–12 SONET Reference Receiver NOTE. In Figure 1–45, the left cursor is at the DC frequency. The right cursor is at the OC-12 data rate of 622 MHz "4 MHz. Also, note that the reference receiver, RR, designation is beside the 20 W vertical scale factor. f. 1–94 Next, move the cursor currently positioned at the beginning of the record until the cursor is just right of the DC spike in the response (see Figure 1–46).
Performance Tests Figure 1–46: Optical impulse response for OC–12 SONET Reference Receiver NOTE. In Figure 1–46 the left cursor has been moved to the zero dB level. The right cursor remains at the OC–12 data rate of 622 MHz "4 MHz (the delta frequency is 584 MHz because the right cursor is no longer referenced to DC). The delta attenuation from the zero dB level to the 622 MHz Cursor is 5.6 dB. 6. Repeat steps 5d through 5f, starting on page 1–93, moving the second cursor to twice the bit rate (1.
Performance Tests Table 1–10: Bessel Thompson frequency response and reference receiver limits Standard Frequency (MHz) Lower Limit (dB) Nominal (dB) Upper Limit (dB) Measured (delta dB) ÁÁÁÁ ÁÁÁÁÁÁ ÁÁÁÁÁÁÁ ÁÁÁÁÁÁ ÁÁÁÁÁÁÁÁ ÁÁÁÁÁÁÁ ÁÁÁÁ ÁÁÁÁÁÁ ÁÁÁÁÁÁÁ ÁÁÁÁÁÁ ÁÁÁÁÁÁÁÁ ÁÁÁÁÁÁÁ ÁÁÁÁ ÁÁÁÁÁÁ ÁÁÁÁÁÁÁ ÁÁÁÁÁÁ ÁÁÁÁÁÁÁÁ ÁÁÁÁÁÁÁ ÁÁÁÁ ÁÁÁÁÁÁ ÁÁÁÁÁÁÁ ÁÁÁÁÁÁ ÁÁÁÁÁÁÁÁ ÁÁÁÁÁÁÁ ÁÁÁÁ ÁÁÁÁÁÁ ÁÁÁÁÁÁÁ ÁÁÁÁÁÁ ÁÁÁÁÁÁÁÁ ÁÁÁÁÁÁÁ ÁÁÁÁ ÁÁÁÁÁÁ ÁÁÁÁÁÁÁ ÁÁÁÁÁÁ ÁÁÁÁÁÁÁÁ ÁÁÁÁÁÁÁ ÁÁÁÁ ÁÁÁÁÁÁ ÁÁÁÁÁÁÁ ÁÁÁÁÁÁ ÁÁÁÁÁÁÁÁ ÁÁÁÁÁÁÁ ÁÁÁÁ ÁÁÁÁÁ
Performance Tests Table 1–10: Bessel Thompson frequency response and reference receiver limits (Cont.
Performance Tests Table 1–10: Bessel Thompson frequency response and reference receiver limits (Cont.
Specifications
Specifications This section begins with a general description of the traits of the TDS 500C, TDS 600B, and TDS 700C oscilloscopes. Three sections follow, one for each of three classes of traits: nominal traits, warranted characteristics, and typical characteristics. Product Description The TDS 500C, TDS 600B and TDS 700C Digitizing Oscilloscopes are portable, four-channel instruments suitable for use in a variety of test and measurement applications and systems. Table 2–1 lists key features.
Specifications Table 2–1: Key features of the TDS 500C, 600B and 700C oscilloscopes (cont.) Feature TDS 600B TDS 500C & TDS 700C Storage Floppy disk drive: 1.44 Mbyte, 3.5 inch, DOS 3.3-or-later floppy disk drive Internal hard disk drive (option HD available on the TDS 500C and 700C) 170 MByte capacity Subject to change due to the fast-moving PC component environment. NVRAM storage for saving waveforms, hardcopies, and setups I/O 1 Full GPIB programmability.
Specifications general purpose knob to adjust some function, such as the position of measurement cursors on screen, or the setting for a channel fine gain. GUI The user interface also makes use of a GUI, or Graphical User Interface, to make setting functions and interpreting the display more intuitive. Some menus and status are displayed using iconic representations of function settings, such as those shown here for full, 250 MHz and 20 MHz bandwidth.
Specifications Table 2–2: Record length and divisions per record vs.
Specifications Trigger System The triggering system supports a varied set of features for triggering the signal-acquisition system. Trigger signals recognized include: H Edge (main- and delayed-trigger systems): This familiar type of triggering is fully configurable for source, slope, coupling, mode (auto or normal), and holdoff. H Logic (main-trigger system): This type of triggering can be based on pattern (asynchronous) or state (synchronous).
Specifications H Use sample, envelope, average and peak detect modes to acquire signals. With the TDS 500C/700C, also use high-resolution mode. H Set the acquisition to stop after a single acquisition (or sequence of acquisitions if acquiring in average or envelope modes) or after a limit condition has been met. H Select channel sources for compliance with limit tests. You can direct the TDS to signal you or generate hard copy output either to a printer or to a floppy-disk file based on the results.
Specifications For time measurements, units can be either seconds or hertz (for 1/time). With the video trigger option installed (Option 05), you can measure the video line number using the vertical cursors. You can measure IRE amplitude (NTSC) using the horizontal cursors with or without the video trigger option installed. Measure Digital Signal Processing (DSP) Measure can automatically extract parameters from the signal input to the digitizing oscilloscope.
Specifications The oscilloscope can also output copies of its display using the hardcopy feature. This feature allows you to output waveforms and other on-screen information to a variety of graphic printers and plotters from the TDS front panel, providing hard copies without requiring you to put the TDS into a system-controller environment. You can make hardcopies in a variety of popular output formats, such as PCX, TIFF, BMP, RLE, EPS, Interleaf, and EPS mono or color.
Nominal Traits This section contains a collection of tables that list the various nominal traits that describe the TDS 500C, TDS 600B, and TDS 700C oscilloscopes. Electrical and mechanical traits are included. Nominal traits are described using simple statements of fact such as “Four, all identical” for the trait “Input Channels, Number of,” rather than in terms of limits that are performance requirements.
Nominal Traits Table 2–4: Nominal traits — Time base system Name Description Range, Sample-Rate1,3 TDS 684B: 5 Samples/sec to 5 GSamples/sec on four channels simultaneously TDS 680B: 5 Samples/sec to 5 GSamples/sec on two channels simultaneously TDS 644B: 5 Samples/sec to 2.5 GSamples/sec on four channels simultaneously TDS 620B: 5 Samples/sec to 2.
Nominal Traits Table 2–4: Nominal traits — Time base system (cont.
Nominal Traits Table 2–5: Nominal traits — Triggering system (cont.) Name Description Ranges, Hold for Ra ges Setup et and a Hol o Feature Min to max ime et /Hol Violation iolatio Trigger igge TimeSetup/Hold Setup Time –100 ns to 100 ns Hold Time –1 ns to 100 ns Setup + Hold Time 2 ns For Setup Time, positive numbers mean a data transition before the clock edge and negative means a transition after the clock edge.
Nominal Traits Table 2–5: Nominal traits — Triggering system (cont.) Name Description Communication Trigger Modes of Operation (Option 2C Comm Trigger) Standard Name Code1 Bit Rate OC1/STM0 OC3/STM1 OC12/STM4 DS0 Sgl DS0 Dbl DS0 Data Contra DS0 Timing E1 E2 E3 E4 E5 (CEPT) STM1E DS1 DS1A DS1C DS2 DS3 DS4NA STS-1 STS-3 FC133 FC266 FC531 FC1063 D2 D1 FDDI NRZ NRZ NRZ Masks2 Masks2 Masks2 Masks2 AMI AMI AMI CMI NRZ CMI AMI AMI AMI AMI AMI CMI AMI CMI NRZ NRZ NRZ NRZ NRZ NRZ NRZ 51.84 Mb/s 155.
Nominal Traits Table 2–6: Nominal traits — Display system (cont.
Nominal Traits Table 2–9: Nominal traits — Mechanical Name Description Cooling Method Forced-air circulation with no air filter. Clearance is required. Refer to the TDS 500C, TDS 600B & TDS 700C User Manual for minimum clearance dimensions. Construction Material Chassis parts constructed of aluminum alloy; front panel constructed of plastic laminate; circuit boards constructed of glass laminate. Cabinet is aluminum and is clad in Tektronix Blue vinyl material.
Nominal Traits 2–16 TDS 500C, TDS 600B & TDS 700C Performance Verification and Specifications
Warranted Characteristics This section lists the various warranted characteristics that describe the TDS 500C, TDS 600B and TDS 700C oscilloscopes. Electrical and environmental characteristics are included. Warranted characteristics are described in terms of quantifiable performance limits which are warranted. NOTE. In these tables, those warranted characteristics that are checked in the procedure Performance Verification appear in boldface type under the column Name.
Warranted Characteristics Table 2–10: Warranted characteristics — Signal acquisition system Name Description Accuracy, DC Gain TDS 600B: ±1.5% for all sensitivities from 2 mV/div to 10 V/div ± 2.0% at 1 mV/div sensitivity TDS 500C, 700C: ±1% for all sensitivities from 1 mV/div to 10 V/div with offset from 0 V to ±100V Accuracy, DC Voltage Measurement, Avera e (using Averaged i Avera Averagee m mode) e) Measurement type DC Accuracy Average of ≥ 16 waveforms TDS 600B: ±((1.
Warranted Characteristics Table 2–10: Warranted characteristics — Signal acquisition system (cont.) Name Description Delay Between Channels, Full Bandwidth TDS 600B: ≤100 ps for any two channels with equal Volts/Div and Coupling settings and both channels’ deskew values set to 0 TDS 500C/700C: ≤50 ps for any two channels with equal Volts/Div and Coupling settings and both channel deskew values set to 0.0 ms. Input Impedance, DC–1 M Coupled 1 M ±0.
Warranted Characteristics Table 2–12: Warranted characteristics — Triggering system Name Description Sensitivity, Edge-Type Trigger, Coupling set e to “DC”1 Trigger source Sensitivity Any Channel TDS 620B & 644B: 0.35 division from DC to 50 MHz, increasing to 1 division at 500 MHz TDS 680B & 684B: 0.35 division from DC to 50 MHz, increasing to 1 division at 1 GHz MHz TDS 500C, 724C, & 754C: 0.35 division from DC to 50 MHz, increasing to 1 division at 500 MHz TDS 784C: 0.
Warranted Characteristics Table 2–13: Warranted characteristics — Output ports, probe compensator, and power requirements (cont.) Name Description Output Voltage and Frequency, Pr be Compensator mpe a r Probe Characteristic Limits Output Voltage 0.
Warranted Characteristics Table 2–14: Warranted characteristics — Environmental Name Description Atmospherics Temperature (no disk in floppy drive): TDS 600B: Operating: +4_ C to +45_ C Floppy disk drive: Operating: +10_ C to +45_ C Nonoperating: –22_ C to +60_ C TDS 500C/700C: Operating: +0_ C to +50_ C Floppy disk drive: Operating: +10_ C to +50_ C Nonoperating: –22_ C to +60_ C Relative humidity (no disk in floppy drive): Operating: 20% to 80%, at or below +32_ C, upper limit derates to 30% relativ
Warranted Characteristics Table 2–14: Warranted characteristics — Environmental (cont.
Warranted Characteristics Table 2–15: Certifications and compliances EC Declaration of Conformity (TDS 500C and TDS 700C) Meets intent of Directive 89/336/EEC for Electromagnetic Compatibility.
Typical Characteristics This subsection contains tables that list the various typical characteristics which describe the TDS 500C, TDS 600B and TDS 700C oscilloscopes. Typical characteristics are described in terms of typical or average performance. Typical characteristics are not warranted.
Typical Characteristics Table 2–16: Typical characteristics — Signal acquisition system (cont.) Name Description Accuracy, Delta Time Measurement The limits are given in the following table for signals having amplitude greater than 5 divisions, reference level = 50%, filter set to (sinX/X), acquired at 5 mV/div or greater. For the TDS 700C, pulse duration < 10 div. Channel skew not included. For the Single Shot condition, 1.
Typical Characteristics Table 2–16: Typical characteristics — Signal acquisition system (cont.) Name Description Effective Bits — TDS 520C & 724C Sample rate The chart on the right gives the typical Input frequency e e ti e bits effective its for o a si sinee wave a e aadjusted ste 1 MHz – 9.2 divs to 9.2 divisions i isio s at 1 MHz, MH 50 mV/div m / i @ 490 MHz – 6.5 divs 25° C. Effective Bits — TDS 540C & 754C 1 GS/s 10 MS/s & HiRes 6.8 bits 9.7 bits 6.
Typical Characteristics Table 2–17: Typical characteristics — Triggering system Name Description Accuracy, Trigger Level or Threshold, DC Trigger source Coupled ple Any Channel (for signals having rise and fall times ≥ 20 ns) Accuracy ±((2% × | Setting – Net Offset |) + (0.3 div × Volts/div Setting ) + Offset Accuracy) Auxiliary Not calibrated or specified Input, Auxiliary Trigger The input resistance is ≥1.5 kW; the maximum safe input voltage is ±20 V (DC + peak AC).
Typical Characteristics Table 2–17: Typical characteristics — Triggering system (cont.) Name Description Width, Minimum Pulse and Rearm, for lse Triggering igge i g Pulse For vertical settings > 10 mV/div. and 3 1 V/div at the BNC input The he mi minimum im m pulse lse widths i ths aand rearm ea m widths i ths and a ttransition a sitio times8 required e i e for o Pulse-Type lse e ttriggering.
Typical Characteristics Table 2–17: Typical characteristics — Triggering system (cont.) Name Description 5 The minimum signal levels required for stable runt pulse triggering of an acquisition. Also, see the footnote for Sensitivity, Edge-Type Trigger, DC Coupled in this table. (Stable counting of events is counting that misses no events). 6 The minimum signal levels required for stable pulse width or glitch triggering of an acquisition.