Datasheet

QPSK Residual EVM (center
frequency = 2 GHz), typical
0.6 % (100 kHz symbol rate)
0.8 % (1 MHz symbol rate)
0.8 % (10 MHz symbol rate)
0.8 % (30 MHz symbol rate)
400 symbols measurement length, 20 Averages, normalization reference = maximum symbol magnitude
256 QAM Residual EVM
(center frequency = 2 GHz),
typical
0.6 % (10 MHz symbol rate)
0.7 % (30 MHz symbol rate)
400 symbols measurement length, 20 Averages, normalization reference = maximum symbol magnitude
LTE Downlink RF measurements
(SV28xx-SVPC)
Standard Supported 3GPP TS 36.141 Version 12.5
Frame Format supported FDD and TDD
Measurements and Displays
Supported
Adjacent Channel Leakage Ratio (ACLR), Spectrum Emission Mask (SEM), Channel Power, Occupied Bandwidth, Power vs. Time
showing Transmitter OFF power for TDD signals and LTE constellation diagram for Primary Synchronization Signal, Secondary
Synchronization Signal with Cell ID, Group ID, Sector ID and Frequency Error.
ACLR with E-UTRA bands
(typical, with noise correction)
1st Adjacent Channel 60 dB (RSA607A)
2nd Adjacent Channel 62 dB (RSA607A)
Mapping (MAPxx-SVPC)
Supported map types Pitney Bowes MapInfo (*.mif), Bitmap (*.bmp), Open Street Maps (.osm)
Saved measurement results Measurement data files (exported results)
Map file used for the
measurements
Google Earth KMZ file
Recallable results files (trace
and setup files)
MapInfo-compatible MIF/MID files
Pulse measurements (SVPxx-
SVPC)
Measurements (nominal) Pulse-Ogram
waterfall display of multiple segmented captures, with amplitude vs time and spectrum of each pulse. Pulse
frequency, Delta Frequency, Average on power, Peak power, Average transmitted power, Pulse width, Rise time, Fall time,
Repetition interval (seconds), Repetition interval (Hz), Duty factor (%), Duty factor (ratio), Ripple (dB), Ripple (%), Droop (dB),
Droop (%), Overshoot (dB), Overshoot (%), Pulse- Ref Pulse frequency difference, Pulse- Ref Pulsephase difference, Pulse- Pulse
frequency difference, Pulse- Pulse phase difference, RMS frequency error, Max frequency error, RMS phase error,Max phase
error, Frequency deviation, Phase deviation, Impulse response (dB),Impulse response (time), Time stamp.
Minimum pulse width for
detection
150 ns
Average ON power at 18 °C to
28 °C, typical
±0.3 dB + absolute amplitude accuracy
For pulses of 300 ns width or greater, duty cycles of .5 to .001, and S/N ratio ≥ 30 dB
Duty factor, typical ±0.2% of reading
For pulses of 450 ns width or greater, duty cycles of .5 to .001, and S/N ratio ≥ 30 dB
Average transmitted power,
typical
±0.5 dB + absolute amplitude accuracy
For pulses of 300 ns width or greater, duty cycles of .5 to .001, and S/N ratio ≥ 30 dB
Peak pulse power, typical ±1.2 dB + absolute amplitude accuracy
For pulses of 300 ns width or greater, duty cycles of .5 to .001, and S/N ratio ≥ 30 dB
Pulse width, typical ±0.25% of reading
For pulses of 450 ns width or greater, duty cycles of .5 to .001, and S/N ratio ≥ 30 dB
Laboratory Spectrum Analyzer
SignalVu-PC applications performance summary
www.tek.com 19