Datasheet
Minimum signal duration for
100% probability of detection
100 μs
Span: 40 MHz, RBW = 300 kHz (Auto)
Due to the non-deterministic execution time of programs running under the Microsoft Windows OS, this specification may not be
met when the host PC is heavily loaded with other processing tasks
Span range (continuous
processing)
1 kHz to 40 MHz
Span range (swept) Up to maximum frequency range of instrument
Dwell time per step 50 ms to 100 s
Trace processing Color-graded bitmap, +Peak, -Peak, average
Trace length 801, 2401, 4001, 10401
RBW range 1 kHz to 4.99 MHz
DPX spectrogram display
Trace detection +Peak, -Peak, Average(V
RMS
)
Trace length, memory depth 801 (60,000 traces)
2401 (20,000 traces)
4001 (12,000 traces)
Time resolution per line 1 ms to 6400 s, user selectable
Spectrum display
Traces Three traces + 1 math trace + 1 trace from spectrogram for spectrum display
Trace functions Normal, Average (VRMS), Max Hold, Min Hold, Average of Logs
Detector Average (VRMS), Average, CISPR peak, +Peak, -Peak, Sample
Spectrum trace length 801, 2401, 4001, 8001,10401, 16001, 32001, and 64001 points
RBW range 10 Hz to 8 MHz
Analog modulation analysis
(standard)
AM demodulation accuracy,
typical
±2%
0 dBm input at center, carrier frequency 1 GHz, 1 kHz/5 kHz input/modulated frequency, 10% to 60% modulation depth
0 dBm input power level, reference level = 10 dBm, Atten=Auto
FM demodulation accuracy,
typical
±1% of span
0 dBm input at center, carrier frequency 1 GHz, 400 Hz/1 kHz input/modulated frequency
0 dBm input power level, reference level = 10 dBm, Atten=Auto
PM demodulation accuracy,
typical
±3% of measurement bandwidth
0 dBm input at center, carrier frequency 1 GHz, 1 kHz/5 kHz input/modulated frequency
0 dBm input power level, reference level = 10 dBm, Atten=Auto
Signal Strength display
Signal strength indicator Located at right side of display
Measurement bandwidth Up to 40 MHz, dependent on span and RBW setting
Tone type Variable frequency based on received signal strength
Laboratory Spectrum Analyzer
SignalVu-PC standard measurements and performance
www.tek.com 15