Datasheet

Exposure
Splash-Proof test, operating
and non-operating
No potential of shock hazard after exposure to non-operating Splash Proof Test per IEC529, level IP52
Dust resistance test, operating
and non-operating
Test method per IEC529, level IP52, test conditions 13.4 and 13.5.
Salt exposure test, structural
parts
Standard MIL-STD-810, Method 509.1, Procedure 1
Dynamics
Vibration
Operating Tektronix Class 2 Random Vibration Test at 2.66 GRMS: 5-500 Hz, 3 Axes at 10 min/axis
Non-Operating MIL-PRF-28800F Class 2
0.030 g
2
/Hz., 10 500 Hz, 30 minutes per axis, 3 axes (90 minutes total)
Shock
Operating Test method per Military Standard MIL-PRF-28800F 1-4
Non-Operating Exceeds the requirements of Military Standard MIL-PRF-28800F
Handling and transit
Bench handling, operating MIL-PRF-28800F Class 2
Transit drop, non-operating MIL-PRF-28800F Class 2
Free-Fall drop, non-operating 32 inches
Spectrum Analyzer
Environmental and safety
www.tek.com 27