Datasheet
Output power level step size
accuracy
± 0.5 dB
Harmonics < -22 dBc, ≥20 MHz
Non-harmonic spurious < -30 dBc; spurious < 2 GHz from TG output frequency
< -25 dBc; spurious ≥ 2 GHz from TG output frequency
Reverse power without
damage
40 Vdc, +20 dBm RF
SignalVu-PC standard measurements and performance
Measurements included.
General signal analysis
Spectrum analyzer Spans from 1 kHz to 18.0 GHz
Three traces plus math and spectrogram trace
Five markers with power, relative power, integrated power,
power density and dBc/Hz functions
DPX Spectrum/Spectrogram Real time display of spectrum with 100% probability of intercept
of up to 15 μsec signals in up to 40 MHz span
Amplitude, frequency, phase vs. time, RF I and Q vs. time Basic vector analysis functions
Time Overview/Navigator Enables easy setting of acquisition and analysis times for deep
analysis in multiple domains
Spectrogram Analyze and re-analyze your signal with a 2-D or 3-D waterfall
display
AM/FM listening Hear, and record to file, FM and AM signals
Analog modulation analysis
AM, FM, PM analysis Measures key AM, FM, PM parameters
RF measurements
Spurious measurement User-defined limit lines and regions provide automatic spectrum
violation testing across the entire range of the instrument. Four
traces can be saved and recalled; CISPR Quasi-Peak and
Average detectors available with option SVQP.
Spectrum emission mask User-defined or standards-specific masks
Occupied Bandwidth Measures 99% power, -xdB down points
Channel Power and ACLR Variable channel and adjacent/alternate channel parameters
MCPR Sophisticated, flexible multi-channel power measurements
CCDF Complementary Cumulative Distribution Function plots the
statistical variations in signal level
SignalVu-PC/RSA507A key
characteristics
Maximum span 40 MHz real-time
9 kHz - 3 GHz swept (RSA503A)
9 kHz - 7.5 GHz swept (RSA507A)
9 kHz - 13.6 GHz swept (RSA513A)
9 kHz - 18.0 GHz swept (RSA518A)
Maximum acquisition time 2.0 s
Spectrum Analyzer
Tracking generator (Option 04)
www.tek.com 17