Amplitude accuracy at all center
frequencies
Center frequency Warranted (18 °C to 28 °C) Typical (95% confidence)
(18 °C to 28 °C)
Typical (-10 °C to 55 °C)
±1.2 dB ±0.8 dB ±1.0 dB
±1.65 dB ±1.0 dB ±1.5 dB
Reference level +20 dBm to -30 dBm, alignment run prior to testing.
Applies to corrected IQ data, with signal to noise ratios > 40 dB.
Accuracy may degrade up to ±0.6 dB after storage at maximum storage temperature, recovers within 24 hours
Intermediate frequency and acquisition system
IF bandwidth 40 MHz
ADC sample rate and bit width 112 Ms/s, 14 bits
Real-time IF acquisition data
(uncorrected)
112 Ms/s, 16-bit integer real samples
40 MHz BW, 28 ±0.25 MHz Digital IF, uncorrected. Corrected values are stored with saved files
Block streaming data at an average rate of 224 MB/s
Block baseband acquisition data
(corrected)
Maximum acquisition time 1 second
Bandwidths ≤ 40 /( 2
N
) MHz, 0 Hz Digital IF, N ≥ 0
Sample rates ≤ 56 / (2
N
) Msps, 32-bit float complex samples, N ≥ 0
Channel amplitude flatness
±1.0 dB, 24 MHz - 6.2 GHz, 18 ⁰C to 28 ⁰C
±1.2 dB, 22 MHz - 24 MHz, 18 ⁰C to 28 ⁰C
±0.5 dB, 24 MHz - 6.2 GHz, -10 ⁰C to 55 ⁰C, typical
±2.5 dB, 22 MHz - 24 MHz, -10 ⁰C to 55 ⁰C, typical
Reference level +20 dBm to -30 dBm, alignment run before testing
Applies to corrected IQ data, with signal to noise ratios > 40 dB
Trigger
Trigger/sync input
Voltage range TTL, 0.0 V – 5.0 V
Trigger level, positive-going
threshold voltage
1.6 V minimum; 2.1 V maximum
Trigger level, negative-going
threshold voltage
1.0 V minimum; 1.35 V maximum
Impedance 10 kΩ
IF power trigger
Threshold range 0 dB to -50 dB from reference level, for trigger levels > 30 dB above the noise floor
Type Rising or falling edge
Trigger re-arm time ≤100 μs
Datasheet
Amplitude
www.tek.com 7
9 kHz -< 3.0 GHz
≥ 3 GHz - 6.2 GHz
Preliminary Datasheet