Datasheet

Digital Phosphor Oscilloscopes
TDS3000C Series
TDS3000C • www.tektronix.com/oscilloscopes
6
Waveform Measurements
Cursors – Amplitude,Time.
Automatic Measurements – Display any four
measurements from any combination of waveforms.
Or display all measurements with measurement
snapshot feature. Measurements include Period,
Frequency, +Width,–Width, Rise time,Fall time,
+Duty cycle, –Duty cycle,+Overshoot, High,Low,
Max, Min, Peak-to-peak,Amplitude,Mean, Cycle
mean, RMS,Cycle RMS, Burst width,Delay,Phase,
Area*
1
, Cycle Area*
1
.
Measurement Statistics – Mean, Min,Max,
Standard deviation. Requires TDS3AAM application
module.
Thresholds – User-definable thresholds for auto-
matic measurements; settable in percent or voltage.
Gating – Isolate a specific occurrence within an
acquisition to take measurements, using either the
screen or cursors.
Waveform Math
Arithmetic – Add, subtract,multiply and
divide waveforms.
FFT – Spectral magnitude. Set FFT vertical scale
to Linear RMS or dBV RMS and FFT window to
Rectangular, Hamming,Hanning or Blackman-Harris.
Advanced Math*
1
Integrate, Differentiate, Define
extensive algebraic expressions including analog
waveforms, math functions,scalars, up to two user-
adjustable variables and results of parametric meas-
urements. For example:(Intg (Ch1–Mean(Ch1)) x
1.414 x VAR1).
Waveform Processing
Autoset – Single-button, automatic setup of all
channels for vertical, horizontal and trigger systems,
with undo autoset.
Deskew – Channel-to-channel deskew +
10 ns may
be manually entered for better timing measurements
and more accurate math waveforms.
Display Characteristics
Display Type 6.5 in. (165.1 mm) liquid crystal
TFT color display.
Display Resolution – 640 horizontal x 480 vertical
pixels (VGA).
Interpolation – Sin(x)/x.
Waveform styles – Dots,vectors, variable persist-
ence, infinite persistence.
Graticules – Full, grid,cross-hair and frame. NTSC,
PAL, SECAM and vectorscope (100% and 75% color
bars) with optional TDS3VID or TDS3SDI application
modules.
Format – YT, XY and Gated XYZ (XY with Z-axis
blanking available on 4-channel instruments only).
Power Source
AC line power
Source voltage – 100 V
RMS
to 240 V
RMS
+10%.
Source frequency – 45 Hz to 440 Hz from 100 V
to 120 V; 45 Hz to 66 Hz from 120 V to 240 V.
Power consumption – 75 W maximum.
Battery power – Requires TDS3BATC, rechargeable
lithium ion battery pack. Operating time,typical –
3 hours.
Environmental and Safety
Operating Non-operating
Temperature 0 °C to +50 °C –40 °C to +71 °C
Humidity Operating and Non-operating: up to 95% RH at or below +30 °C
Operating and Non-operating: up to 45% RH +30 °C up to +50 °C
Altitude To 3,000 m 15,000 m
Electromagnetic Compatibility Meets or exceeds EN61326 Class A,Annex D radiated and conducted emissions and immunity;
EN6100-3-2 AC Powerline Harmonic Emissions; EN6100-3-3 Voltage Changes,Fluctuation and Flicker;
FCC 47 CFR, Part 15, Subpart B, Class A; Australian EMC framework
Safety UL61010B-1, CSA1010.1, IEC61010-1,EN61010-1
Physical Characteristics
Instrument
Dimensions mm in.
Width 375 14.8
Height 176 6.9
Depth 149 5.9
Weight kg lbs.
Instrument only 3.2 7
with accessories 4.5 9.8
Instrument Shipping
Package Dimensions mm in.
Width 502 19.8
Height 375 14.8
Depth 369 14.5
Rackmount (RM3000)
Dimension mm in.
Width 484 19
Height 178 7
Depth 152 6
*
1
Requires TDS3AAM application module.