Performance Verification
AWG510 & AWG520 Service Manual
4-51
!!
%
"# #"
%# % $"
Figure 4-19: Trigger Signal (-5V check1)
d. Push Cursor, %, &,
, keys as the low level of a pulse to be set to
–5.35V.
e. Verify that a sine wave is displayed on the oscilloscope.
!!
%
Figure 4-20: Trigger Signal (-5V check2)
6. Push the RUN button to turn off the RUN LED.
7. Disconnect all the cable.