User Manual
Remote operation
Harmonics configuration (cont.)
Syntax
:HMX:VLT:RN
G <value>
:HMX:AMP:RNG <value>
:HMX:WAT:RNG <value>
Where <valu
e> is the maximum harmonic to be display in the range of
1 to 100
Descripti
on
If harmonics measurements are selected with the :SEL command (See
page 71.), the power analyzer will display all the harmonics up to the
number sp
ecified by <value>. The displayed harmonics can be restricted
to only odd numbered harmonics using the harmonic sequence command.
This command works on a group. Use the :INST:NSEL command first
to s e lect
the active group.
Syntax
:HMX:VLT:FOR <value>
:HMX:AMP:FOR <value>
:HMX:WAT:FOR <value>
Where <value> is 0 for absolute values or 1 for percentage values
Description
If harmonics measurements are selected with the :SEL command (See
page 71.), the power analyzer can display all harmonics (except the first) as
an absolute value or as a percentage of the fundamental (first) harmonic.
This command works on a group. Use the :INST:NSEL command first
to select the active group.
:HMX
:VLT/AMP:DF
Commands for setting up the distortion factor measurements.
Distortion factor setup
Syntax
:HMX:VLT:DF:REF <val ue>
:HMX:AMP:DF:REF <val ue>
Where <value> is 0 for fundamental or 1 for rms
Description
For distortion factor readings (also known as the difference formula),
the reference on the denominator of the equation can be either the rms
reading or the fundamental harmonic reading.
This command works on a group. Use the :INST:NSEL command first
to select the active group.
PA3000 Power Analyzer User Manual 75