User Manual
Remote operation
:SEL
Select results
Syntax
:SEL:ALL
:SEL:ALL:GRP<group>
:SEL:CLR
:SEL:CLR:GRP
<group>
:SEL:<measurement>
Where <group> is a group number 1 through 4.
Where <measu
rement> is:
VLT – Volts rms
AMP – Amps rms
WA T – Watts
VAS – VA
VAR – VAr
FRQ – Frequ
ency
PWF – Power factor
VPK+ – Volts peak (positive)
VPK- – Volt
s peak (negative)
APK+ – Amps peak (positive)
APK- – Amps peak (negative)
VDC – Volt
sDC
ADC – Amps DC
VRMN – Volts rectified mean
ARMN – Amp
srectified mean
ACMN – Amps corrected rectified mean
VCMN – Volts corrected rectified mean
VCF – Vol
ts crest factor
ACF – Amps crest factor
VTHD – Volts Total Harmonic Distortion
VDF – Vo
lts Distortion Factor
VTIF – Volts Te lephone Influence Factor
ATHD – Amps Total Harmonic Distortion
ADF – Am
ps Distortion Factor
ATIF – Amps Telephone Influence Factor
IMP – Impedance
RES – R
esistance
REA – Reactance
HR – Integrator time
1
WHR – W
att H ours
1
VAH – VA Hours
1
VRH – VAr Hours
1
AHR –
Amp Hours
1
WAV – Average Watts
1
PFAV – Average Power Factor
1
CVAR
– Correction VArs
1
VAHF – Fundamental VA-hours
VARHF – Fundamental VAr-hours
VF –
Fundamental Volts rms
AF – Fundamental Amps rms
WF – F undamental Watts
PA3000 Power Analyzer User Manual 71