User Manual
Remote operation
Interfacing w
ith GPIB systems (optional)
The PA3000 optionally supports control v ia a GPIB port. This option must be
installed by an authorized Tektronix representative.
See IEEE 488/GPIB for a detailed pin description of the GPIB connector. (See
page 148, I EEE 488 / GPIB (Optional).)
Status reporting
Status byte
The PA3000 u ses a similar status byte to IEEE 488.2. The PA3000 Status Byte
Register
(STB) contains the ESB and DAS bits. These two bits indicate a
non-zero state in the Standard Event Status Register (ESR) or the Display Data
Status Register (DSR) respectively.
The ESR and DSR each have enable registers, ESE and DSE respectively, set by
the user. These enable registers act as a mask to reflect chosen elements of the
appropriate status registers to the Status Byte Register. Setting the appropriate bit
of the enable register to 1 configures the bits to be summarized in the STB.
If a status byte is read, the DSR and ESR registers are cleared.
F
igure 31: Status byte
PA3000 Power Analyzer User Manual 63