User Manual
The menu system
Update rate
The update rate
determines how often new results are available in the instrument.
The values in the Update Rate menu list indicate the time interval (in seconds)
between e very update of the results. Restrictions are placed on the number and
type of results permitted at faster update rates.
The range is 0.2 s to 2 s in 0.1 s increments, where 0.5 is the default value.
With update rates below 0.5 seconds, the number of results that can be updated
at that rate is limited.
Autozero
Autozero is a method of automatically canceling any small parasitic signals, such
as DC offsets, in the measurement. Three options are available:
On. (default) The instrument runs autozero every minute.
Off. When autozero is disabled the instrument uses the last autozero values.
Run Now.
The instrument runs a n immediate autozero on the currently
selected ranges. This takes approximately 100 ms. The state of whether
autozero is either enabled or disabled will not be changed, and there is no
feedback to indicate it has run.
Clock
Use the following options to check or set the internal clock:
Set Time. Enter the time using the format shown and pr ess to confirm.
Set Date. Enter the date using the format shown and press to confirm.
Time Format. Select 12 Hour or 24 Hour and press to confirm.
Dat
eFormat. Select the required date format and press
to confirm.
Power saving
The instrument has the ability to reduce its own power consumption by switching
off the display.
In the Display menu you have the following options:
Always On. This is the default mode and the display will always be on.
Switch off after 10 minutes. The display turns off after 10 minutes if no key is
pressed. Pressing any key turns the display on again.
Switch off in remote mode. If the instrument receives a command through
any of the communication interfaces, the display turns off. Pressing any key
turns the display on again, but the instrument will remain in the remote mode
until the LOCAL key is pressed. PressingtheLOCALkeytoswitchthe
display on w ill not return the instrument to local mode.
60 PA3000 Power Analyzer User Manual