User Manual

The menu system
VA r H o u r s
Amp Hours
Watts Average
PF Average
Volts
Amps
Watts
Fundamental VA-Hours (VAHf)
Fundamental VAr-Hours (VArHf)
Correction VArs
For each waveform selected, there is a choice on the graph menu to turn on or off
the selected parameter for each channel in the group.
Integrator graphing parameters are set on a per-group basis. To change groups use
the left and right arrow keys at the bottom left of the display.
Additional information is available for setting up the integrator. (See page 43,
Integrator mode.) Additional information is available for displaying the integrator
waveforms. (See page 17, Integrator screen.)
Interfaces
This menu can be used to set up the remote control interfaces of the PA3000.
RS-232 baud rate
9600, 19200, and 38400 (default) are available.
The PA3000 uses hardware handshaking (RT
S / CTS) with no parity, eight data
bits and one stop bit (N,8,1).
The RS-232 baud rate is unchanged after a “*RST” or “:DVC” command.
GPIB address
Enter the GPIB address.
Default address is 6. The address is unchanged after a “*RST” or “:DVC”
command.
Ethernet
The PA3000 offers Ethernet communications
through an Ethernet port u sing
TCP/IP.
The Ethernet port will make a TCP/IP connection on port 5025. Port 5025 is
designated by the Internet Assigned Numbers Authority (IANA) to be a SCPI port.
54 PA3000 Power Analyzer User Manual