The menu system
After selectin
g Ballast mode you need to set up the fundamental frequency at
which power will be transmitted. This is typically 50, 60, or 400 Hz. The setup
screen can b e found under the Modes Setup Modes Ballast Setup.The
power analyzer uses this to adjust the measurement window to t the specied
frequency.
The returned frequency is not the fundamental power frequency; but the ballast
switching frequency. This is also the frequency used for harmonic analysis.
When Ballast mode is selected, the frequency range is set to >10 Hz” and the
bandwidth is set to “High” for the group. These setting are locked out in Ballast
mode and, upon the return to normal mode, the settings are restored.
Standby P
ower mode
Standby Power mode integrates the watts, amps, VA and PF readings over
a user-specied measurement period. This is a requirement of many standby
power standards.
Driven by consumer demand and energy efciency legislation, there is an
ever-increasing need to measure power consumption of products while they are
in stan
dby mode.
One of the most widely used standards for measurement is IEC 62301. Part of
this s
tandard requires the measurement of power over a prolonged period of time
without missing any short duration power events. The Standby Power mode
provides continuous sampling of voltage and current to produce a n accurate watts
measurement over the user specied period.
In Standby Power mode, you must specify the integration window in seconds.
Watts, amps, PF, and VA will then be integrated over the specied period. All
other results will be updated at the normal user specied update rate.
The integration period depends on the combination of the specied window and
theupdaterateoftheinstrument.(Seepage60,Update rate.) This is because the
results will be integrated over a n exact multiple of the update rate. For example, if
t
he update rate is 0.5 seconds (default), then the integration period will always be
exactly as specied. However, if an update rate of 0.4 seconds is requested, then
integration period will switch between 1.2 seconds and 0.8 seconds.
For the most accurate measurements it is recommended that the ranges are xed
during the measurement period. (See page 50, Fixed/Auto ranging.)
Integrator mode
Integrator mode provides measurements for determining energy consumption by
integrating measurements over a specied period or by running continuously.
Integration can be started manually, by a threshold trigger, or at a specicvalue.
In addition, for certain parameters, average values are also a vailable.
PA3000 Power Analyzer User Manual 43