User Manual
The menu system
The reference v
alue can be either the fundamental reading or the RMS reading.
The default reference is the fundamental value.
Total Harmonic Distortion. THD is a measure of the distortion of a waveform.
Under the V and A measurement menus is the ability to set the following
parameters:
Harmonic Reference. Fundamental or RMS (default Fundamental).
Harmonic Sequence. Odd and Even or Odd Only (default is Odd and Even).
Harmonic R
ange. 2 to 100 (default 7). This is the last harmonic used in the
calculation. If Odd Only harmonics are specified and the Range is set to an
even number, the preceding harmonic will be the last one used.
Harmonic Zero. Exclude or Include (default Exclude)
For the d
istortion settings and harmonics setting, the values will be remembered
whether the actual displayed reading is turned on or off. For example, if the
number of harmonics to be displayed is changed from 7 to 13, turning off and then
on again the display of voltage harmonics will not affect this setting.
The formula for voltage and current THD are:
and
The Total Harmonic Distortion formula (previously called the series formula)
will produce more accurate results for harmonic noise when the THD is less
than 5%. When selecting the THD formula, it is important to set the maximum
harmonics setting to an appropriately large number to get valid results. The higher
the harmonic count the more accurate the calculation.
Telephone influence factor. The telephone influencefactor(TIF)isaTHD
measurement weighted at frequencies within the bandwidth of a normal telephone
circuit. It is a measure of how the voltage or current distortion in electrical power
circuits might interfere with adjacent telephone circuits. TIF measurements a re a
requirement of sta ndards such as ANSI C50.13 "Rotating Electrical Machinery -
Cylindrical-R
otor Synchronous Generators" and are most often used o n standby
power generators and UPS. The harmonics included in a TIF measurement are
from1to73oddandeven.
The formula for voltage and current TIF are:
Default refe
rence = Fundamental
40 PA3000 Power Analyzer User Manual