User Manual
Getting started
The instrument
will start the power on seque nce; this takes approximately
15 seconds. During power on you will see the serial number and firmware
version of the instrument.
4. The instrument is now ready for us e.
Concept of global, group, and channel parameters
Definition of a group
With a multiphase power analyzer there is often a requirement to link together
measurement channels. This is known as grouping. Within a group, one
channel will act as the frequency source a nd reference for all other channels
in the group. Grouping is commonly used in applications such as three phase
motor measurements. Channels 1 and 2 can be grouped together to allow for
the m easurement of the input power, where Channels 3 and 4 could be grouped
together to measure the output power. For more information on applying grouping
to channels, see the Wiring section of The Menu System chapter. (See page 47,
Wiring.)
Global, group, and channel
settings
The PA3000 has many different settings that affect both the appearance of the
results and the actual results. To make the instrument easier to operate, settings
might have an effect on one or more parameters. Depe nding on the parameter, the
influence or use of it might be on a global level, a per-group level or a per-channel
level. The split for parameters that have an effect on measurements and results
is defined below.
Global settings
Global settings affect all measurements. The following settings are global:
B
lanking (See page 59, Blanking.)
Averaging (See page 59, Averaging.)
Update rate (See page 60, Update rate.)
Autozero (See page 60, Autozero.)
Global settings will appear under the System Configuration menu.
Groups settings
Per-group settings affect every channel in a group. The settings affected are:
Measurements (See page 36, Measurements.)
Measurement configuration(Seepage38,Measur ement Configuration menu.)
Mode (See page 42, Modes.)
Wiring (See page 47, Wiring.)
Ranges (See page 50, Ranging.)
Shunt selection (See page 51, Shunts.)
PA3000 Power Analyzer User Manual 3