Getting started
Before each use
Ensure that the
power and connecting cables as well as all accessories and
connected devices used in conjunction with this product are in proper working
order and clean.
Ensure that any third-party accessories used in conjunction with the device
conform to the applicable IEC 61010-031 / IEC 61010-2-032 standards and
are suitable for the respective measuring voltage range.
Connection sequ ence
WARNING. To avoid possible electric shock or personal injury, when the
measuring
circuit is used to measure MAINS, the voltage to earth must not exceed
600 V
rms
in a CAT II environment.
For safety reasons, when connecting a circuit to the power analyzer, proceed
in the sequence outlined as follows:
1. Connect the instrument power cord to a properly grounded mains outlet.
The power analyzer is now connected to the protective earth ground wire.
2. Power on the instrument.
3. Conne
ct the measuring circuit according to all instructions and as shown in
the connection diagrams in this manual.
During use
For connecting cables and instruments, work in teams of at least two people.
If you detect any damage to the housing, controls, power cable, connecting
leads, or connected devices, immediately disconnect the unit from the power
sup
ply.
If you are in doubt as regards the safe operation of the device, immediately
sh
ut down the unit and the respective accessories, secure them against
inadvertent switching on, and have them serviced by a qualied service
person.
Power on
1. Check that the power analyzer is in good condition with no signs of damage.
2. Follow the Connection Sequence described in the Before you begin - safety
section. (See page 1.)
3. Press the power switch to turn on the power analyzer.