User Manual

Application examples
Figure 58: Inrush current measurement wiring diagram
Method 1: Inrush current
measurements (PA3000
direct)
The following steps describe the process for setting up the PA3000 for the inrush
current measurements.
1. Set the PA3000 to the default conguration:
a. Press
.
b. Scroll down to User Conguration and press
.
c. Select Load Default Conguration, and then press
to conrm.
The PA
3000 will load the defaults and display a conrmation scre en. Press
and then return to the Main Menu by pressing .
NOTE
. The PA3000 needs to be congured to continuously sample and return
non-averaged results. It also needs to be congured to stor e peak values.
2. Ran
ging must be xed on both the voltage and current channels to enable
continous sampling. To select a manual range, go to the Main Menu (press
), go to Inputs Ranging Current/Voltage range, and then select
an appropriate range.
If the expected inrush current is not known, select the highest range and then
dial down, as required, after repeating the test as needed for more accurate
r
esults
134 PA3000 Power Analyzer User Manual