Application examples
The test will di
splay the Pass/Fail status based on all evalua ted parameters.
The test summary can be reviewed under the Results tab. All of the required
parameters wi
ll be summarized with a Pass/Fail status under Test Summary and
General Results.
The Power Re
adings tab allows you to scroll through the time scale and to debug
specic issues on a given time stamp. Use the scroll bar on the top of the window
to scroll through the time line.
The test results can be exported as a full report by clicking the Full Report PDF
icon on the top ribbon. Alternately, the raw data can exported using the Export
CSV icon.
Figure 57: IEC 62301 Ed. 2.0 Standby power test report
Further settings if required
Ranging. By default, the PWRVIEW software sets up the PA3000 current range
in Auto-Up-Only. The Auto-Up-Only option ranges up the PA3000 until it nds
the highest range where the input signal is not clipped. This helps with setting
the best range for the test. Alternatively, if the expected peak currents are known,
a manual range can be selected for higher accuracy. The manual range can be
selected for the selected shunt through the Setup page under Range options.
132 PA3000 Power Analyzer User Manual