User Manual
Application examples
Method 1: Energy
consumption
measurements (PA3000
direct)
The following s
teps describe the process to set up energy consumption testing on
home or office appliances using the PA3000.
Figure 51: Energy consumption testing on the PA3000
1. Set the PA3000 to the default configuration:
a. Press
.
b. Scroll down to User Configuration and press
.
c. Select Load Default Configuration, and then press
to confirm.
The PA3000 will load the defaults and display a confirmation screen. Press
and then return to the Main Menu by pressing .
2. To enable the Integrator mode, go to Modes → Select Mode → Integrator.
3. Press
, select Setup Modes → Integrator Setup, and select the desired
start method.
Manual works by pressing the INTEG RUN key to start and stop
integration.
Clock allows for setting up a specific time to start integration.
L
evel allows you to set trigger levels on a speci fi c signal to start the
integration measurement.
PA3000 Power Analyzer User Manual 123