User Manual
Remote operation
:SYST:POWER
Power usage
Syntax
:SYST:POWER:DISP <va lue>
Where <value> is 0, 1, or 2
Description
This command allows the display to be switched off to reduce the power
consumption of the power analyzer. The operation of the display is
determined by the following values:
0–alwayson
1 – off after 10 minutes with no key press nor remote control
2 – off in remote control m ode
Syntax
:SYST:POWER:DISP?
Returns 0–alwayson
1 – off after 10 minutes with no key press nor remote control
2 – off in remote control m ode
User configuration commands
These commands relate to the User Configuration menu item.
:CFG:USER
User configurations
Syntax
:CFG:U
SER:LOAD <value>
:CFG:USER:SAVE <valu e>
Where <value> is user configuration 1 through 8 for saving and 0 through
8forlo
ading. 0 is the default configuration.
Returns
1fors
uccess or 0 for failure
Description
Thes
e commands will be used to load and save one of the user
configurations.
Syntax
:CFG
:USER:REN <value>,<c onfig.Name>
Where <value> is user configuration 1 through 8 and <config.Name> is a
new configuration name (up to 16 characters)
Description
This command changes the name of the configuration to aid in finding
it a
gain.
Tip: When saving or loading configurations, allow at least three seconds
after sending the :CFG:USER:LOAD <value> command before reading a 1
fo
r success, or 0 for failure.
PA3000 Power Analyzer User Manual 95