User Manual
Remote operation
Interface com
mands
Interface commands are used to set up and control the various ways of
communicating with the power analyzer.
:COM:RS2
RS-232 configuration
Syntax
:COM:RS2:BAUD <baud rate>
Where <baud rate> is a baud rate of 9600, 19200, or 38400
Description
This command sets the RS-232 baud rate.
Syntax
:COM:RS2:BAUD?
Returns
Baud rate of 9600, 19200 or 38400
:COM:IEE
GPIB configuration
Syntax
:COM:IEE:ADDR <addre ss>
Where <address> is an address in the range of 1 to 30
Description
This command sets the GPIB address for the PA3000.
Syntax
:COM:IEE:ADDR?
Returns
The GPIB address for the power analyzer. If -1 is returned, no GPIB card
is installed.
:COM:ETH
Return Ethernet configurations
Syntax
:COM:ETH:SUB?
:COM:ETH:IP?
:COM:ETH:GATE?
SUB – Subnet mask
IP – IP address
GATE – D efault gateway
Returns
Number in the form of v4 IP address xxx.xxx.xxx.xxx
Description
This command returns the requested information in the form of an IP
address. The information returned is the current configuration. If DHCP is
used as the assignment method, then the values returned would be those
values assigned by the DHCP server.
:COM:ETH:STAT
Static Ethernet configuration
Syntax
:COM:ETH:STAT <value>
Where <value> is 0 or 1
Description This command determines whether the power analyzer uses a static IP
address or one assigned by a DHCP server. If <value> = 0 then a DHCP
server is used. If <value> = 1 then the static IP settings are used.
PA3000 Power Analyzer User Manual 89