User Manual

Remote operation
Scaling (cont.)
Returns
A number from 0.00001 to 100000
Description This command is linked to a channel, not a group. It returns the scaling
factor for the currently selected channel. Use the :INST:NSELC command
rst to select the active channel.
:ANA
Analog inputs
Syntax
:ANA <input>,<range >
Where <input> is an input number 1 through 4 and <range> = 1 or 10
Description
This command sets the analog inputs 1 though 4. If <range> is 1, then the
±1 V range selected. If <range> is 10, then the ±10 V range is selected
for the specied input.
Syntax
:ANA? <input>
Where <input> is an input number 1 through 4.
Returns The measured analog signal on the selected input
Graph a
nd waveform commands
:WAV
Waveform graphs
Syntax
:WAV:VLT <enable>
:WAV:AMP <enable>
:WAV:WAT <enable>
Description
These commands enable or disable the display of a waveform graph,
where <enable> is 1 to enable or 0 to disable. The commands are all
channel based commands. Select the channel with :INST:NSELC before
using these commands.
Syntax
:WAV:VLT?
:WAV:AMP?
:WAV:WAT?
Description
These commands return the enable or disable status of the waveform
graph display. The commands are all channel based commands. Select
the channel with :INST:NSELC before using these commands.
Returns
1 if the waveform is enabled or 0 if the waveform is not enabled.
88 PA3000 Power Analyzer User Manual