User Manual
Remote operation
Total harmonic distortion setup (cont.)
Syntax
:HMX:VLT:TH
D:NZ <value>
:HMX:AMP:THD:NZ <val ue>
Where <value> is 0 for exclude or 1 for include
Description
For total harmonic distortion (THD) readings, the formula can either include
or exclude
the DC component.
This command works on a group. Use the :INST:NSEL command first
to select the active group.
:HMX:VLT/AMP:TIF
Telephone influence factor setup
Syntax
:HMX:VLT:TIF:REF <va lue>
:HMX:AMP:TIF:REF <va lue>
Where <value> is 0 for fundamental or 1 for rms
Description
For telephone influence factor readings, the reference on the denominator
of the equation can be either the rms reading or the fundamental harmonic
reading.
This command works on a group. Use the :INST:NSEL command first
to select the active group.
:MIN
Minimum column
Syntax
:MIN <value>
Where <value> is 0 for disabled or 1 for enabled
Description The MIN command adds a column to the results that displays the minimum
value of each parameter since the last time the minimum values were
reset. A column is added for each channel in the group, as well as for
sum results, if they are selected.
Enabling the column will always reset the MIN and MAX values for the
currently selected group. The values can also be reset by using the :RES
command or pressing the R ESET/CLE AR key on the front panel.
To reset the MIN hold values, send the command :MIN 1 to re-enable the
column. Note that both MIN and MAX hold values will be reset.
This command works on a group. Use the :INST:NSEL command first
to select the active group.
Syntax
:MIN?
Returns 0 or 1
Description
This command returns the status of the minimum value column. A 0 will be
returned if disabled; a 1 if enabled.
This command works on a group. Use the :INST:NSEL command first
to select the active group.
PA3000 Power Analyzer User Manual 77