Datasheet
Supplemental characteristics
The following specifications are supplemental characteristics that provide additional information about instrument functions and performance. These characteristics are non-
warranted specifications; they describe the typical performance of the PA3000.
For detailed information on the PA3000 Power Analyzer product specifications refer to the Model PA3000 Power Analyzer Instrument Specifications document, Tektronix part
number 077-1252-00 downloadable from www.tek.com/manual/downloads.
Hours accuracy ±0.0125%
Watts accuracy (45 Hz to 850 Hz) ±(V
rms
acc × A
rms
) ±(A
rms
acc × V
rms
)
Watt hour accuracy ±(Wattacc + Houracc)
VA accuracy (45 Hz to 850 Hz) ±(V
rms
acc × A
rms
) ±(A
rms
acc × V
rms
)
PF accuracy Wacc / VA
Frequency accuracy
0.1 Hz to 10 Hz 0.1% of reading
10 Hz to 1 MHz 0.5% of reading
Analog inputs
Ranges 10 V
dc
range: ±1 V to ±10 V
V
dc
range: ±0.1 V to ±1 V
Accuracy ±0.2% of reading ±0.2% of range ±0.005 V
Sample rate 1000 samples per second
Communications
IEEE-488 (Option) IEEE Std 488.1 compliant
RS-232 Baud rates from 9600 bps, 19200 bps (default), and 38400 bps
8 bit, No parity, 1 stop bit, hardware flow control
9 pin male D-type connector
Ethernet IEEE 802.3 compatible, 10Base-T
RJ-45 connector with Link and Activity indicators
TCP/IP connection on port 5025
USB device USB 2.0 compatible, Full speed (12 Mb/sec)
Datasheet
14 www.tek.com