Datasheet
Specifications
All specifications are guaranteed unless noted otherwise. All specifications apply to all models unless noted otherwise. Specifications are subject to change without notice.
For detailed information on the PA3000 Power Analyzer product specifications refer to the Model PA3000 Power Analyzer Instrument Specifications document, Tektronix part
number 077-1252-00 downloadable from www.tek.com/manual/downloads.
Current accuracy
Accuracy A
rms
(45 Hz to 850 Hz)
1 A shunt ±0.04% reading ±0.04% range
30 A shunt ±0.04% reading ±0.04% range
External shunt ±0.1% reading ±0.04% range
Accuracy A
rms
, typical (10 Hz to 45 Hz, 850 Hz to 1 MHz)
1 A shunt ±(0.05 + 2×10
-5
× f)% reading ±0.05% range ±40 μA
30 A shunt ±(0.05 + 2×10
-5
× f)% reading ±0.05% range ±4 mA
External shunt ±(0.1 + 2×10
-5
× f)% reading ±0.05% range ±1.1 mV
Accuracy A
dc
1 A shunt ±0.05% reading ±0.1% range ±100 μA
30 A shunt ±0.05% reading ±0.1% range ± 10mA
External A shunt ±0.1% reading ±0.1% range ±1.1 mV
Ranges (peak)
1 A shunt 12.5 mA, 25 mA, 50 mA, 125 mA, 250 mA, 500 mA, 1.25 A, 2.5 A, 5 A
30 A shunt 500 mA, 1 A, 2 A, 5 A, 10 A, 20 A, 50 A, 100 A, 200 A
External A shunt 50 mV, 100 mV, 200 mV, 500 mV, 1 V, 2 V, 5 V, 10 V, 20 V
Voltage accuracy
Accuracy V
rms
(45 Hz to 850 Hz) ±0.04% reading ±0.04% range
Accuracy V
rms
(10 Hz to 45 Hz,
850 Hz to 1 MHz), typical
±(0.05 + 1×10
-5
× f)% reading ±0.05% range ±20 mV
Accuracy V
dc
±0.05% reading ±0.1% Range ±50 mV
Ranges (peak) 5 V, 10 V, 20 V, 50 V, 100 V, 200 V, 500 V, 1000 V, 2000 V
PA3000 Power Analyzer