Adjustment Proc
edures
Equipment Required for Adjustment Procedure
In addition to the equipment required to perform the performance verication, the adjustment procedures require the
equipment listed in Table 7.
Table 9: Test equipment
Description Minimum requirements Example product
Digitizing oscilloscope (Required for
adjustment procedure only)
Bandwidth 100 MHz, average acquisition
mode, vertical sensitivity 2 mV/div
Tektronix TDS3000 series
Function generator (Square and sine
wave output)
Output level adjustable to 10 Vpk-pk,
Separate Trigger or Sync output
Tektronix AGF300 series
50 termination (needed only if
oscilloscope does not support 50
termination)
50 ±1
011-0045-xx
P6250 & P6251 Technical Reference 43