User Manual
Adjustment Proc
edures
Equipment Required for Adjustment Procedure
In addition to the equipment required to perform the performance verification, the adjustment procedures require the
equipment listed in Table 7.
Table 9: Test equipment
Description Minimum requirements Example product
Digitizing oscilloscope (Required for
adjustment procedure only)
Bandwidth ≥100 MHz, average acquisition
mode, vertical sensitivity 2 mV/div
Tektronix TDS3000 series
Function generator (Square and sine
wave output)
Output level adjustable to 10 Vpk-pk,
Separate Trigger or Sync output
Tektronix AGF300 series
50 Ω termination (needed only if
oscilloscope does not support 50 Ω
termination)
50 Ω ±1 Ω
011-0045-xx
P6250 & P6251 Technical Reference 43