User Manual
Performance Ver
ification
Verification.
4. Disconnect the BNC cable from the BNC connector in the probe socket on the 1103 power supply.
5. Connect the BNC cable to the BNC-to-probe tip adapter.
6. Connect the probe into the probe socket on the 1103 power supply.
7. Connect the probe tip to the DM pins of the BNC-to-probe tip adapter.
8. Set the probe to 4.25 V range (÷ 5), DC reject off, full bandwidth.
9. Adjust the reference level of the spectrum analyzer to display a signal on screen.
10. Record the level displayed on the spectrum analyzer.
40 P6250 & P6251 Technical Reference