Analog Bandwidth
Normalize the Setup
1. Set the synthesizer for 500 MHz/10 dBm for P6250 probe and 1 GHz/10 dBm for the P6251 probe.
2. Set the spectrum analyzer:
Center frequency at 500 MHz for the P6250 probe, and 1 GHz for the P6251 probe.
Span to 10 MHz.
Resolution bandwidth to auto.
Reference level to 10 dBm.
Vertical sensitivity to 2 dB /div.
3. Record the level displayed on the spectrum analyzer. (This level represents the synthesizer output minus the signal
path loss.)
P6250 & P6251 Technical Reference 39