User Manual

Performance Ver
ication
14. Calculate the attenuation twice, u sing the values from the measurements as follows: (V
in
1-V
in
2) ÷ (V
out
1-V
out
2)
and (V
in
2-V
in
3
)÷(V
out
2-V
out
3).
15. Verify that the
attenuation is in the range of 49 to 51. Record the results in the test record.
4.25 V Range 5).
16. S et the input voltage on the DC source to approximately 4.2 V ±50 mV. Record the actual voltage as V
in
1.
17. Change the probe voltage range to 4.25 V 5).
18. Record the output voltage as V
out
1.
19. S et the input voltage on the DC source to 2 V, and record the actual voltage as V
in
2.
20. Record the output voltage as V
out
2.
21. Set the input voltage on the DC source to approximately 0.5 V. Record the actual voltage as V
in
3.
22. Record the output voltage as V
out
3.
23. Calculate the attenuation twice, using the formulas from the previous test. Refer to step 14.
24. Verify that the attenuation is in the range of 4.9 to 5.1. Record the results in the test record.
25. S lide the DC Reject switch on the probe to turn DC reject on.
26. Verify that the output returns to approximately 0 volts. (This is a functional check; there is no specied performance limit.)
27. S lide the DC Reject switch on the probe to turn DC reject off.
28. K eep the output connections for the next procedure.
P6250 & P6251 Technical Reference 21