User Manual
Performance Ver
ification
Preparation
CAUTION. To prevent damage to the 80E04 sampling head, plug in the sampling head and then power on the TDS8000
oscilloscope
.
Prepare the equipment as follows:
1. Connect the 80E04 sampling head to channel 1 of the sampling oscilloscope and then power on the TDS8000
oscilloscope.
2. Plug in and power on all test equipment. (See Table 5 on page 16.)
3. Connect the probe to channel 1 of the 1103 power supply.
4. Allow the probe and test equipment to warm up for 20 minutes at an ambient temperature of 20 °C to 30 °C.
5. Photocopy the test record and use it to record the test results. (See Table 7 on page 35.)
Perform the verification procedures in order.
P6250 & P6251 Technical Reference 19