User Manual
Warranted Chara
cteristics
Warranted Cha
racteristics
Warranted characteristics describe guaranteed performance wit hin tolerance limits or certain type-tested requirements. (See
Table 1.) Warranted characteristics that have checks in the Performance Veri fication section are marked with the
symbol.
Table 1: Warranted electrical characteristics
Specification P6250 P6251
Differential signal range (DC c oupled)
±4.25 V
(DC + peak AC)
;3V
RMS
±42 V
(DC + peak AC)
;30V
RMS
±4.25 V
(DC + peak AC)
;3V
RMS
±42 V
(DC + peak AC)
;30V
RMS
Common-mode signal range (DC coupled)
±35V
(DC + peak AC)
;25V
RMS
±35V
(DC + peak AC)
;25V
RMS
Maximum nondestructive input voltage
between signal and common of the same
channel
±100 V
(DC + peak AC)
±100 V
(DC + peak AC)
DC attenuatio
n accuracy
±5% of input (both ranges) ±5% of input (both ranges)
Bandwidth (4.25 V range (÷ 5), probe
only)
DC to ≥500 MHz DC to ≥1GHz
Rise time, 4.25 V range (÷ 5), probe
only (10–90%,
+20°Cto+30°C)
≤700 ps ≤350 ps
Common-mode re jec tion ratio
42 V range (÷ 5
0):
>55 dB at 30 kHz
>50dBat1MHz
>18 dB at 250 MH
z
42 V range (÷ 5
0):
>55dBat30kHz
>50 dB at 1 MHz
>18 dB at 250 MH
z
Temperature
Operating: 0
to +40 °C (32 °F to +104 °F)
Nonoperating: –55 to +75 °C (–67 °F to –167 °F)
1
Humidity
Operating: 0-90% RH, tested at + 30 to +40 °C
Nonoperating: 0-90% RH, tested at +30 to +60 °C
1
See warning that follows.
WARNING. To avoid a burn hazard at high ambient temperatures, do not touch the probe with bare hands at nonoperating
temperatures above +50 °C .