Service Information
NOTE. The R10 ad
justment affects the aberrations of both the 1mA/mV and
10 mA/mV termination setting along with the high frequency response. Readjust
the high frequency response if it fails.
Figure 8: Location of probe and termination adjustments
Mainten
ance
The information in this section will help you maintain your probe for a long
service life.
Clean
ing
To clean the probe body, use a soft cloth dampened with water. To clean the core,
open the jaw and clean the exposed core surfaces with a cotton swab dampened
with isopropyl alcohol.
CAUT
ION. To prevent damage to probe materials, avoid using chemicals that
contain benzene, benzine, toluene, xylene, acetone, or similar solvents.
Do not immerse the probe in liquids or use abrasive cleaners.
Do not lubricate the jaws mating surfaces. Any lubricant between the core pieces
should be removed with a recommended solvent.
Do not use a petroleum based lubricant on the plastic. If the plastic slide assembly
re quires lubrication, use a silicone-based grease sparingly.
When cleaning the probe, look for any excessive wear of the slide parts that might
cause the probe to operate improperly. Dirty or worn mating surfaces between
the transformer and the lid can degrade low-frequency response. Clean these
surfaces if necessary.
24 P6022 Instruction Manual