Service Information
1. Set the P6022 te
rmination sensitivity to 1 mA/mV.
2. Connect the calibration fi xture to the Leveled Sinewave Output of the
calibration g
enerator.
3. Connect the P6022 termination and probe to CH1 of the oscilloscope.
4. Set the oscilloscope Vertical Deflectionto20mV/div.
5. Set the oscilloscope Horizontal Scale to 10 μs/div.
6. Set the oscilloscope Auto Measurement to display "Amplitude".
7. Set the Leveled Sinewave Output of the calibration g enerator to 50 kHz and
an amplitude of 5 V
pp
.
8. Connect the probe to the calibration fixture.
9. Enable t
he Leveled Sinewave Output of the calibration generator.
10. Measure and record the oscilloscope "amplitude" Auto Measurement as M1 in
the tab
le. (See Table 8 on page 20.)
11. Set the oscilloscope Horizontal Scale to 4 ns/div or 5 ns/div.
12. Set the Leveled S inewave O utput of the calibration generator to 100 MHz.
13. Measure and record the oscilloscope amplitude Auto Measurement as M2 in
the table.
14. Calculate the high frequency bandwidth ratio:
P6022 Instruction Manual 19