User manual
Service Information
1. Set the P6022 te
rmination sensitivity to 1 mA/mV.
2. Connect the Leveled Sinewave Output of the calibration generator to a BNC
"T". Connect o
ne branch of the BNC "T" to the calibration fixture. Connect
the other branch of the BNC "T" to the DMM input using a BNC-to-dual
banana adapter.
3. Set the DMM to ACV (autorange).
4. Set the Leve
led Sinewave Output of the calibration generator to 50 kHz and
an amplitude of 5 V
pp
.
5. Measure an
d record the DMM output as M1.
6. Disconnect the coax cable from the BNC-to-dual banana adapter.
7. Connect the P6022 termination and probe to the BNC-to-dual banana adapter.
8. Connect the probe to the calibration fixture.
9. Record the DMM output as M2.
10. Calculate the % of error.
11. Rec ord the results in the test record and compare the results against the
midband ac curacy specification. (See Table 1 on page 9.)
12. Set the P6022 termination sensitivity to 10 mA/mV.
13. Record the DMM output as M3.
14. Calculate the % of error:
15. Record the results in the test record and compare the results against the
midband accuracy specification.
16 P6022 Instruction Manual