User manual

Service Information
Table 6: Equipm
ent list
Item
Characteristic
Description Recommended eq
uipment
Bandwidth
DC to 300 MHz
Vertical sen
sitivity
1 m V/div
Oscilloscope
Measurement functions
Amplitude a
veraging
TDS 303X, or TDS 305X
Fast rise ti
me
Rise time 1
ns
1V
P-P
into 50
Wavetek 910
0withoption
100:250 or Tektronix PG 506A
Calibration generator
Sinewave voltage 5 V
P-P
into 50 (100 mA),
935 Hz to 120 MHz
1.5X, at
ness
Wavetek 91
00 with option
100:250, or Tektronix SG
5030, SG 503
Digital m
ultimeter (DMM)
RMS ACV 5 1/2 digi
ts or better
50 kHz 0.5%
Keithley 2000, or HP 3458A
Adapter
BNC "T"
Tektron
ix part number
103-0030-XX
Coaxial cables (2) 36-inch, 50 precision
Tektron
ix part number
012-0482-XX
Adapte
r
BNC-to-dual banana
Tektro
nix part number
103-0090-XX
Fixtur
e
Current probe, calibration
Tektro
nix part number
067-0559-XX
Photocopy the Test Record and use it to record the performance test results. (See
Table 9 on page 21.)
14 P6022 Instruction Manual