Instructions
Service Information
P6021 Instruction M anual
23
C13
R10
R12
TEKTRONIX, INC
125 TURNS
P6021 AC CURRENT PROBE
Probe (back)
Figure 8: Location of probe adjustments
R36
C25C22
Figure 9: Location of termination adjustments
8. Set the termination sensitivity control to the 10 mA/mV position.
9. Set the oscilloscope Vertical Deflection to 1 mV/division.
10. Readjust adjustments as needed to ensure the optimum pulse response on
both sensitivity settings.