Adjustments
P6015A Instruction Manual
2-5
Long-Form Procedure
The long-form com pensation procedure typically should be
necessary only under these conditions:
H the probe head or compensation box has been replaced
H the probe has come out of compensation because of aging effects
over a prolonged period
Access to adjustments in this procedure requires removal of the top
half of the compensation box (Figure 2-- 1). When reinstalling the
top, note that the e dges are asymetrical and tha t the top will seat
securely only whe n it is c orrectly orie nted. This ensures t hat the
adjustment holes are properly aligned with the circuit board.
Figure 2- 1: Access to Long-Form Adjustments