User manual

Adjustments
2-4
P6015A Instruction Manual
Table 2- 1: Test Equipment Required for Long-Form Adjustment
Item
Minimum
Requirements
Recommended
Example Application
Oscilloscope
Input Impedance:1 MΩ
Sensitivity: 1 mV/div
Bandwidth: 100 MHz
Tektronix 11402A
with 11A32 plug-
in, or Tektronix
TDS 460
All checks and
adjustments
Calibration
Generator
Rise time: 10 ns
Repetition rate: 1 MHz
Amplitude: 50 V
Tektronix
PG 506A
1,2
All checks and
adjustments other
than bandwidth
Leveled Sine
Wave Generator
Range: 50 kHz to 75 MHz
Amplitude: 5V
Tektronix SG 503
2
Bandwidth check
BNC Male-to-GR
Adapter
Tektronix
part number
017-0064-00
All checks and
adjustments
BNC 50 Ω
Termination
Tektronix
part number
011-0049-01
Transient
response check
and adjustment,
bandwidth check
1
Oscilloscopes with less than 1 mV/div sensitivity can be used while
making mo st checks and adjustments. However, b ecause of the 1000X
probe attenuation, the system will not display sufficient deflection for
optimum adjustment of transient response unless a calibration generator
having higher amplitude output is substituted.
2
Requires TM 500 or TM 5000 Series Power Module or equivalent.