User manual

Specifications
P6015A Instruction Manual
1-27
Typical and Nominal Characteristics
This section lists the various typical and nominal characteristics that
describe the P6015A High Voltage Probe.
Nominal characteristics are determined by design and/or inspection.
Nominal characteristics do not have tolerance limits.
Typical characteristics are described in terms of typical or average
performance. Typi cal characteristics are not warranted.
Table 1- 5: Typical Electrical Characteristics
Characteristic Information
Input resistance
100 MΩ ±2%. See Figure 1--8 for typical input
impedance curve.
Input capacitance 3 pF when probe is properly LF compensated. See
Figure 1--8 for typical input impedance curve.
LF compensation range 7pFto49pF
Aberrations 25% p-p for the first 200 ns on a 100 MHz oscilloscope
when used with 10 in (25.4 cm) ground lead.
<10% p-p typical after first 200 ns; ±5% after the first
400 ns.
Temperature coefficient of DC
attenuation
0.006% per degree C
1
Voltage coefficient of DC attenu -
ation
0.018% per kV
Delay time 10 ft cable: 14.7 ns
25 ft cable: 33.3 ns
1
Resistor temperature rose 60_Cat20kVrmsovera30minuteperiod.