User manual

Specifications
P6015A Instruction Manual
1-25
Table 1- 4: Warranted Environmental Characteristics
Characteristic Information
Temperature
Nonoperating
-- 5 5 _Cto+75_ C(--67_ F to +167_F)
Operating
DC + peak AC
Peak Pulse
0_Cto+35_C(+32_Fto+95_F)
0_Cto+50_C(+32_F to +122_F)
(See Table 1-- 1 on page 1--10 and Time Limitations
Specification below )
Humidity
Nonoperating / Operating 95% relative humidity at +50 °C (+122 °F). See
Figure 1--7 for derating characteristics.
Maximum altitude
Nonoperating 15,000 m (50,000 ft)
Operating 4,600 m (15,000 ft)
Peak pulse voltage derated from 40 kV at 8000 feet
(2440 m) to 30 kV at 15,000 feet (4570 m).
Vibration (random)
Nonoperating 3.48 g rms from 5 to 500 Hz. Ten minutes on each axis.
Operating 2.66 g rms from 5 to 500 Hz. Ten minutes on each axis.
Shock (nonoperating) 500 g, half sine, 0.5 ms duration, 18 shocks total in three
axis.
Time Limitations
Less than 70% of Rated
Input Voltage at 0--35_C
No time limit
Greater than 70% of
Rated Input Voltage at
0--35_C
30 minutes maximum
in any 2.5 hour period
35--50_C
15 minutes maximum
in any 2.5 hour period