P6015A Instruction Manual
Table 1- 3: Warranted Electrical Characteristics
Characteristic Information
Maximum input voltage
DC + peak AC
1
1.5 kV to 20 kV. See frequency derating curve in
Figure 1--4. (DC plus peak AC rating is limited to
temperatures below 35° C.)
Peak pulse 40 kV
a
(Never exceed 20 kV rms)
Duty cycle derating -- 100 ms maximum duration at 10%
maximum duty cycle. See duration and duty cycle
derating curve in Figure 1--5.
Altitude derating -- Peak pulse derated linearly from
40 kV at 8000 feet (2440 m) to 30 kV at 15,000 feet
(4570 m) altitude.
Relative Humidity (RH) derating -- Voltage derated with
increasing temperature and relative humidity (see
Figure 1--7).
Bandwidth (-- 3 dB) Test conditions: Test oscilloscope bandwidth must be
100 MHz, Z
source
=25Ω
10-ft cable
75 MHz
25-ft cable
25 MHz
Rise Time
2
10-ft cable
4.67 ns (calculated from bandwidth)
25-ft cable
14 ns (calculated from bandwidth)
DC attenuation 1000:1 ±3% (Excluding
oscilloscope error)
Test conditions: Oscillo-
scope input resistance
mustbe1MΩ ±2%
1
Characteristic not checked in manual
2
T
r