P6015A Instruction Manual
Low-Frequency Compensation
1. Set the calibration generator to high amplitude output, and set the
period to 1 ms. Set the pulse amplitude to display five divisions.
2. Set the oscilloscope time base to display 200 s/div.
3. Center the waveform on the screen.
4. Adjust LF COMP (C5) so that the leading corner of the square
wave is level with the trailing corner.
5. Adjust MID 1 (C1) to flatten the area 200 s from the leading
corner. Refer to Figure 1 --6 to determine the zones affected by
this and the following adjustments.
50 s
Mid 3
200 s
Mid 1
100 s
Mid 2
Figure 1- 6: Zones Affected by Compensation Adjustments