Instructions

The large circuit board visible inside the instrument is the Main board.
10. Disconnect the power cable from the Main board by pressing on the tab near the wire-end of the connector and sliding the connector
away from the board.
11. Disconnect the display cable from the Main board by flipping up the latch on the backside of the connector and sliding the cable out.
Use the illustration printed on the Main board near the connector for reference.
12. Disconnect the touch screen cable from the Main board by flipping up the latch on the cable-side of the connector and sliding the
cable out of the connector.
13. Disconnect the front panel cable from the Main board by flipping up the latch on the cable-side of the connector and sliding the cable
out of the connector. This cable includes hook features, so the cable must be lifted out of the connector while sliding it out.
How to sanitize an instrument
2 Series Mixed Signal Oscilloscope MSO22, MSO24 Declassification and Security Instructions 9